FDA Medical Device Recalls (openFDA)
Terumo Cardiovascular Systems Corporation: Sarns TCM II, Catalog Numbers: 4415; 4416;164925; 164930;164935;164940 UDI 00886799000014 The Sarns TCM is indicated for controlling and monitoring patient temperature .
- Recall number
- Z-1418-2018
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Sarns TCM II, Catalog Numbers: 4415; 4416;164925; 164930;164935;164940 UDI 00886799000014 The Sarns TCM is indicated for controlling and monitoring patient temperature .
- Status
- Terminated
- Initiated
- 2018-03-16
- Posted
- 2018-04-18
- Terminated
- 2020-06-10
- Reason
- Update to cooler-heater cleaning instructions.
- Root cause
- Labeling design
- Action
- The firm,Terumo Cardiovascular Systems Corporation, mailed customers an "Urgent Medical Device Correction" letter dated 2/16/18. The letter described the product, problem and actions to be taken. The Customers were instructed to do the following: 1) discard the current Sarns TCM, TCM II Cooling and Heating, or tHX2 Temperature Management Systems cleaning guides; 2) Review the notice and updated cleaning guides, and ensure all users receive notice of the updated cleaning guides and; 3) Return the Customer Response Form acknowledging receipt of the notice via Terumo Recall Email: tcvs.recall@terumomedical.com or Terumo Recall Fax: 734-741-6149. The letter also includes a list of the updates to the cleaning guides. Customers with questions may contact Terumo CVS Customer Service (800) 521-2818, Monday-Friday, 8am-6pm ET or Fax (734)741-6149.
- Quantity
- 2393- total for all products
- Distribution
- Worldwide Distribution: US (Nationwide), Australia, Belgium, Canada, Chile, China. Colombia, Dominican Republic, England, Finland, France, Germany, Greece, HK, India, Indonesia, Iran. Israel, Italy, Japan, Korea, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates,and Vietnam.
- Lot, serial or model codes
- All Lot Codes
- 510(k) numbers
- ["K883603"]
- PMA numbers
- not on file
- Product code
- DWC
- Firm FEI
- 1828100
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28