FDA Medical Device Recalls (openFDA)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: 2D Perfusion, release R1.0.x, R1.1.x, R1.2, and R1.2.1
- Recall number
- Z-1417-2023
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- 2D Perfusion, release R1.0.x, R1.1.x, R1.2, and R1.2.1
- Status
- Open, Classified
- Initiated
- 2023-02-20
- Posted
- 2023-04-20
- Terminated
- not on file
- Reason
- There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
- Root cause
- Software design
- Action
- A recall notification letter was issued February 20, 2023. The Urgent Medical Device Correction letter includes the following instructions for the Customer: " Stop using SmartPerfusion and 2D Perfusion. NOTE: This message does not impact (other parts of) the Philips Allura or Azurion systems or other tools within the Interventional Workspot. The Philips Allura and Azurion systems as well as other Interventional Workspot tools may continue to be used. " Place the Urgent Medical Device Correction Letter with the documentation of the system. " Circulate the notice to all users so that they are aware of the product issue. " Return the reply form attached to the Urgent Medical Device Correction letter to Philips to confirm that the users of the system have reviewed and understood it.
- Quantity
- 549 units (75 US, 474 OUS)
- Distribution
- Worldwide - US Nationwide distribution.
- Lot, serial or model codes
- UDI: (01)00884838059276 2D Perfusion with Allura Xper or Azurion model number: 722003, 722006, 722008, 722010, 722012, 722013, 722020, 722023, 722026, 722027 722028, 722029, 722035, 722038, 722039, 722058, 722059, 722064, 722067 722068, 722078, 722079, 722223, 722224, 722225, 722226, 722227 and 722228. Software License Keys: 459801071511 sw key 2D Perfusion (IW) 459800240601 sw key 2D Perfusion (XV)
- 510(k) numbers
- ["K132147"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 3002807880
- Firm city
- Best
- Firm state
- not on file
- Firm country
- Netherlands
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28