Skip to the content

FDA Medical Device Recalls (openFDA)

Merge Healthcare, Inc.: Merge Cardio software using EchoIMS. The firm name on the label is Merge Healthcare, Hartland, WI.

Recall number
Z-1415-2017
Recalling firm
Merge Healthcare, Inc.
Product
Merge Cardio software using EchoIMS. The firm name on the label is Merge Healthcare, Hartland, WI.
Status
Terminated
Initiated
2016-04-04
Posted
2017-03-08
Terminated
2017-03-20
Reason
A situation can occur allowing two physicians to access the same study report in EchoIMS when launched from the Cardio Study List without receiving the read-only notification prompt.
Root cause
Software design
Action
Merge sent an Urgent: Medical Device Recall letter dated March 28, 2016, via e-mail on April 4, 2016. A letter was issued to nonresponders on July 11, 2016. The letter notifies the customer of the issue, informs them of the workaround, and that no further action was being taken to correct the issue, therefore, the customer needs to be aware of the possible behavior. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they understand the workaround. The response was required no later than April 18, 2016. Nonresponders were issued a second letter dated July 11, 2016, asking for a response by August 2, 2016. The letter and response form contained the same information as the first letter. If the second letter results in additional unresponsive customers, a call campaign will be managed by Customer Support and documented via SalesForce. Three call attempts will made before closing the case. For further information please call (877) 741-5369.
Quantity
17 sites potentially have the affected versions
Distribution
US Distribution was made to medical facilities in CA, FL, IL, MD, MO, OH, OK, TX, VT, and WI. Military distribution was also made.
Lot, serial or model codes
Version 9.0.9 and earlier using EchoIMS.
510(k) numbers
["K051649"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000116130
Firm city
Hartland
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register