Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Co.: EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Product Usage: The EpiCenter system is a clinical data management system designed for use with certain BD instrument systems. The EpiCenter system provides advanced capabilities in data management, including extended demographic data, comprehensive reporting features including a powerful database query tool, bi-directional Laboratory Information System (LIS) communications and long-term data storage. Additionally, the EpiCenter system provides an Expert System that can manipulate and/or act upon AST results.

Recall number
Z-1414-2017
Recalling firm
Becton Dickinson & Co.
Product
EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System . Product Usage: The EpiCenter system is a clinical data management system designed for use with certain BD instrument systems. The EpiCenter system provides advanced capabilities in data management, including extended demographic data, comprehensive reporting features including a powerful database query tool, bi-directional Laboratory Information System (LIS) communications and long-term data storage. Additionally, the EpiCenter system provides an Expert System that can manipulate and/or act upon AST results.
Status
Terminated
Initiated
2014-08-25
Posted
2017-02-27
Terminated
2017-03-08
Reason
The information in the EpiCenter interface translation tables was not updated to match the Bruker MALDI BioTyper database identifications for some organisms. This could lead to a possible mis-identification of an organism.
Root cause
Vendor change control
Action
Customers were initially contacted via phone calls and subsequently provided with a written communication. For questions contact BD Technical Services Department at 1-800-638-8663.
Quantity
Qty Distr. to Field - 115
Distribution
15 US customers, 18 Ex-US sites, and 0 distributors. Approximately 115 customers totally and 0 distributors
Lot, serial or model codes
Lot Numbers Version 6.20A  Cat. Number 442953
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register