FDA Medical Device Recalls (openFDA)

Medtronic Neurological: Medtronic Intrathecal Catheter, Model 8731, (part of infusion systems using Medtronic implantable pumps) The catheter is designed for use in the intrathecal space. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA.

Recall number
Z-1414-06
Recalling firm
Medtronic Neurological
Product
Medtronic Intrathecal Catheter, Model 8731, (part of infusion systems using Medtronic implantable pumps) The catheter is designed for use in the intrathecal space. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA.
Status
Terminated
Initiated
2006-07-21
Posted
2006-09-09
Terminated
2007-06-21
Reason
Tip dislodgement during implantation-Medtronic is recalling Model 8731 Intrathecal Catheter and Model 8598 Intrathecal Catheter Distal Revision Kit because the platinum-iridium tip may be dislodged by the guide wire during implantation.
Root cause
Other
Action
An Urgent Worldwide Medical Device Recall letter was sent to affected customers 07/20/06 for delivery on 07/21/06. The letter describes the issue, identifies the affected product and list action required by the customer. Affected devices are to be removed from active inventory, quarantined and returned to Medtronic for a warranty credit. A Reply card is asked to be returned. Medtronic Representatives will follow-up by July 28, 2006.
Quantity
Recalls Z-1414-1415-06-USA - 359 devices. Foreign - 575 devices
Distribution
Worldwide Distribution-USA and Austria, Belgium, Canada, Cyprus, Denmark, Finland, France, Germany, Greece, India, Ireland, Italy, Netherlands, Norway, Poland, Portugal, San Marino, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, and the United Kingdom
Lot, serial or model codes
Model 8731 Base Lot Numbers: B011311N, B011421N, B011422N, B011500N, B011721N, B011737N, B011738N, B011786N, N0012097, N0012140, N0012141, N0012521, N0012692, N0012814, N0012815, N0012816, N0013149, N0013150, N0013352, N0013353, N0013354, N0013355, N0013707, N0013710, N0013936, N0014007, N0014236, N0014237, N0014330, N0014331, N0014332, N0014722, N0015050.   *Model 8731 Catheters and Model 8598 Distal Revision kits have a Base Lot number of eight (8) characters plus a two (2) digit suffix that represents serialization within these recalled lots. This list provides just the Base lot Number (the first eight characters of the product Lot No. found on the product labeling).
510(k) numbers
not on file
PMA numbers
["P860004"]
Product code
LKK
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register