Encore Medical, Lp: RSP Impaction Fixture

Recall number
Z-1413-2017
Recalling firm
Encore Medical, Lp
Product
RSP Impaction Fixture
Status
Terminated
Initiated
2017-02-09
Posted
not on file
Terminated
2017-10-20
Reason
During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.
Root cause
Device Design
Action
There are two field safety notices, one for Consignees who have surgeons that use the impaction fixtures (Verson 1) and one for Consignees who had previously indicated their surgeons do NOT use the impaction fixtures (Verson 2). The two recall notification letters were sent out on 2/9/17.
Quantity
626 units
Distribution
US, South Korea, Australia, Canada, United Kingdom/Ireland, Germany
Lot, serial or model codes
109931L01, 109931L02, 115670L15, 115670L16, 128092L08, 128092L09, 137917L16, 167829L06, 52258L01, 52258L01A, 67428L01, 76386L01, 76386L02, 81722L01, 81722L02
510(k) numbers
not on file
PMA numbers
not on file
Product code
HWA
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register