Encore Medical, Lp: RSP Impaction Fixture
- Recall number
- Z-1413-2017
- Recalling firm
- Encore Medical, Lp
- Product
- RSP Impaction Fixture
- Status
- Terminated
- Initiated
- 2017-02-09
- Posted
- not on file
- Terminated
- 2017-10-20
- Reason
- During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.
- Root cause
- Device Design
- Action
- There are two field safety notices, one for Consignees who have surgeons that use the impaction fixtures (Verson 1) and one for Consignees who had previously indicated their surgeons do NOT use the impaction fixtures (Verson 2). The two recall notification letters were sent out on 2/9/17.
- Quantity
- 626 units
- Distribution
- US, South Korea, Australia, Canada, United Kingdom/Ireland, Germany
- Lot, serial or model codes
- 109931L01, 109931L02, 115670L15, 115670L16, 128092L08, 128092L09, 137917L16, 167829L06, 52258L01, 52258L01A, 67428L01, 76386L01, 76386L02, 81722L01, 81722L02
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HWA
- Firm FEI
- 1000116912
- Firm city
- Austin
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28