Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Medline Convenience Kits: 1) ADMIT KIT, Model Number: DYKA1343A

Recall number
Z-1410-2026
Recalling firm
Medline Industries, LP
Product
Medline Convenience Kits: 1) ADMIT KIT, Model Number: DYKA1343A
Status
Open, Classified
Initiated
2026-01-07
Posted
2026-02-17
Terminated
not on file
Reason
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Root cause
Under Investigation by firm
Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers regarding Proxima Sterile Surgical Gowns, Packs, and Drapes. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-26-003 Recall Code: 3. Your account will receive credit once the response form is submitted. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers regarding Medline Kits containing Proxima Sterile Surgical Gowns and Drapes. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 Recall Code: 3. Upon receipt of your submitted response form,
Quantity
1070 units
Distribution
Worldwide distribution - US Nationwide.
Lot, serial or model codes
1) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 25LBC732; 2) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 25FBN781; 3) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 25DBK346; 4) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 24IBD021; 5) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 24DBK171; 6) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 24BBV821; 7) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 23HBP415; 8) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 23EBC552; 9) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 23CBE825; 10) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 22LBR618; 11) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 22KBA245; 12) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 22EBK467; 13) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 22ABE640; 14) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 21JBF397; 15) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 21IBE395; 16) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 21IBB933; 17) DYKA1343A, UDI-DI: 10193489312966(each), 40193489312967(case), Lot Number: 21CBV232
510(k) numbers
not on file
PMA numbers
not on file
Product code
NSB
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register