Blease Medical Equipment, Ltd.: The BleaseSirius and BleaseFocus Anaesthesia Systems.

Recall number
Z-1408-2010
Recalling firm
Blease Medical Equipment, Ltd.
Product
The BleaseSirius and BleaseFocus Anaesthesia Systems.
Status
Terminated
Initiated
2010-03-26
Posted
2010-04-16
Terminated
2011-08-23
Reason
Potential for delivery of a lower than set concentration of anaesthetic to a patient due to a sticking valve. The lower shuttle may become jammed, not permitting flow of anaesthesia gas through the vapouriser. The vapouriser back bar mounting valves will require exchange only on the systems identified in this matter, no other systems are affected.
Root cause
Other
Action
On 2/26/2010 notification letters were sent to all international consignees. The 2 domestic consignees were sent notification letters via FEDEX on 3/26/2010. The domestic consignees were also contacted by phone on 3/26/2010. The notification letter is titled URGENT MEDICAL DEVICE CORRECTION. It advises users of the issue and provides additional information which would permit the continued use of the Blease system until replacement back bar valves can be fitted by the recalling firm's service representatives. For users who choose to continue to use the device, the manufacturer recommends the use of an Agent Monitor and user education; also checks that vapourisers have been mounted correctly and the system is leak free, and testing in advance of a procedure. For additional questions, US Customers are directed to call Global Technical Support - +1 425-657-7200 ext: 5089. Non US Customers, call Global Technical Support - +44 1494 784422.
Quantity
40 units
Distribution
Distributed to 2 U.S. hospitals and to customers in the following countries: Panama, Poland, Australia, Singapore, United Kingdom, Bangladesh, Nicaragua, Vietnam, Argentina, Finland, and Columbia.
Lot, serial or model codes
Serial numbers:  Siri 000116, Siri 000885, Siri 001100, Siri 001609, Siri 001610, Siri-001618, Siri-001626, Siri-001637, Siri-001640, Siri-001644, Siri-001658, Siri-001666, Siri-001667, Siri-001668, Siri-001674, AND Siri-001681.  Focu-000735, Focu-000737, Focu-000738, Focu-000739, Focu-000740, Focu-000741, Focu-000742, Focu-000743, Focu-000744, Focu-000745, Focu-000746, Focu-000747, Focu-000748, Focu-000749, Focu-000750, Focu-000751, Focu-100001, Focu-100003, Focu-100004, Focu-100005, Focu-100006, Focu-100007, Focu-100008, AND Focu-100009.
510(k) numbers
["K051629"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
3002806541
Firm city
Chesham Bucks
Firm state
England
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register