FDA Medical Device Recalls (openFDA)
ErgoSafe Products, LLC (DBA) Prism Medical: A-series lift - A625 Manual Part No. 324117. Prism Medical.
- Recall number
- Z-1406-2017
- Recalling firm
- ErgoSafe Products, LLC (DBA) Prism Medical
- Product
- A-series lift - A625 Manual Part No. 324117. Prism Medical.
- Status
- Terminated
- Initiated
- 2017-01-30
- Posted
- 2017-02-28
- Terminated
- 2017-12-04
- Reason
- During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely.
- Root cause
- Device Design
- Action
- The recalling firm called each customer on Jan 30, 2017 to review the recall and schedule a time for the firm to come to their facility to personally swap the recalled product for a new device. The FSNs were sent on Jan 31, 2017.
- Quantity
- 76
- Distribution
- Product was distributed to MT and NV; one VA consignee.
- Lot, serial or model codes
- A6MC000016 A6MC000019 A6MC000020 A6MC000021 A6MC000022 A6MC000023 A6MC000024 A6MC000025 A6MC000026 A6MC000027 A6MC000028 A6MC000029 A6MC000030 A6MC000031 A6MC000009 A6MC000002 A6MC000004 A6MC000001 A6MC000003 A6MC000005 A6MC000006 A6MC000007 A6MC000008 A6MC000010 A6MC000011 A6MC000012 A6MC000013 A6MC000014 A6MC000015 A6MC000017 A6MC000018 A6MC000096 A6MC000087 A6MC000088 A6MC000090 A6MC000091 A6MC000092 A6MC000093 A6MC000094 A6MC000048 A6MC000068 A6MC000070 A6MC000072 A6MC000073 A6MC000074 A6MC000075 A6MC000078 A6MC000079 A6MC000081 A6MC000082 A6MC000083 A6MC000085 A6MC000086 A6MC000042 A6MC000043 A6MC000045 A6MC000046 A6MC000047 A6MC000049 A6MC000050 A6MC000051 A6MC000052 A6MC000053 A6MC000054 A6MC000055 A6MC000056 A6MC000057 A6MC000058 A6MC000059 A6MC000060 A6MC000061 A6MC000062 A6MC000063 A6MC000064 A6MC000066 A6MC000071
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FSA
- Firm FEI
- 3007802293
- Firm city
- Maryland Heights
- Firm state
- MO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28