Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: PrecisePLAN 2.10 Treatment Planning System

Recall number
Z-1406-2010
Recalling firm
Elekta, Inc.
Product
PrecisePLAN 2.10 Treatment Planning System
Status
Terminated
Initiated
2006-03-15
Posted
2010-04-16
Terminated
2010-12-30
Reason
Error in dose rates which were calculated at exactly twice the expected values.
Root cause
Software design
Action
As described in a prior related submission, Phase I - "Important Notice A285, Scrolling Selected Frame Dose Calculator Error" dated March 15, 2006 was distributed to affected customers to inform them of the potential and what steps should be taken. Phase 2 -- Software release PrecisePLAN 2.11 was released July 11, 2006 to correct this problem. All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. Phase 3 -- All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. The company is directing questions to the local Electa representative or to their offices in the U.K. at telephone 44 (0)-1293-654200.
Quantity
40 units
Distribution
Worldwide: IL, NY, PA, TN, Canada, China, Czech Republic, Germany, India, Israel, Italy, Mexico, Poland, Russia, Spain, Switzerland
Lot, serial or model codes
1762685762, 1762688017, 1762688506, 1726288507, 1762676911, 1762688175, 1762689384, 1762688019, 1762688022, 1762688876, 1762689442, 1762688177, 1762690813, 1762686657, 1762689588, 1762689386, 1762684357, 1762686833, 1762684370, 1762678238, 1762689519, 1762684359, 1762690814, 1762688514, 1762578223, 1762678239, 1762684360, 1762684361, 1762684362, 1762684367, 1762683737, 1762682414, 1762688081, 1762686584, 1762686843, 1762686832, 1762688024, 1762688080, 1762686836, 1762684358
510(k) numbers
["K022411"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register