FDA Medical Device Recalls (openFDA)
Elekta, Inc.: PrecisePLAN 2.10 Treatment Planning System
- Recall number
- Z-1406-2010
- Recalling firm
- Elekta, Inc.
- Product
- PrecisePLAN 2.10 Treatment Planning System
- Status
- Terminated
- Initiated
- 2006-03-15
- Posted
- 2010-04-16
- Terminated
- 2010-12-30
- Reason
- Error in dose rates which were calculated at exactly twice the expected values.
- Root cause
- Software design
- Action
- As described in a prior related submission, Phase I - "Important Notice A285, Scrolling Selected Frame Dose Calculator Error" dated March 15, 2006 was distributed to affected customers to inform them of the potential and what steps should be taken. Phase 2 -- Software release PrecisePLAN 2.11 was released July 11, 2006 to correct this problem. All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. Phase 3 -- All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. The company is directing questions to the local Electa representative or to their offices in the U.K. at telephone 44 (0)-1293-654200.
- Quantity
- 40 units
- Distribution
- Worldwide: IL, NY, PA, TN, Canada, China, Czech Republic, Germany, India, Israel, Italy, Mexico, Poland, Russia, Spain, Switzerland
- Lot, serial or model codes
- 1762685762, 1762688017, 1762688506, 1726288507, 1762676911, 1762688175, 1762689384, 1762688019, 1762688022, 1762688876, 1762689442, 1762688177, 1762690813, 1762686657, 1762689588, 1762689386, 1762684357, 1762686833, 1762684370, 1762678238, 1762689519, 1762684359, 1762690814, 1762688514, 1762578223, 1762678239, 1762684360, 1762684361, 1762684362, 1762684367, 1762683737, 1762682414, 1762688081, 1762686584, 1762686843, 1762686832, 1762688024, 1762688080, 1762686836, 1762684358
- 510(k) numbers
- ["K022411"]
- PMA numbers
- not on file
- Product code
- MUJ
- Firm FEI
- 1037831
- Firm city
- Norcross
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28