Skip to the content

FDA Medical Device Recalls (openFDA)

ErgoSafe Products, LLC (DBA) Prism Medical: A-series lift - A625 Manual w/Load Cell & Wi-Fi, Omni Part No. 324124. Prism Medical.

Recall number
Z-1405-2017
Recalling firm
ErgoSafe Products, LLC (DBA) Prism Medical
Product
A-series lift - A625 Manual w/Load Cell & Wi-Fi, Omni Part No. 324124. Prism Medical.
Status
Terminated
Initiated
2017-01-30
Posted
2017-02-28
Terminated
2017-12-04
Reason
During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely.
Root cause
Device Design
Action
The recalling firm called each customer on Jan 30, 2017 to review the recall and schedule a time for the firm to come to their facility to personally swap the recalled product for a new device. The FSNs were sent on Jan 31, 2017.
Quantity
21
Distribution
Product was distributed to MT and NV; one VA consignee.
Lot, serial or model codes
A6MC000110 A6M0001010 A6M0001011 A6M0001012 A6M0001013 A6M0001014 A6M0001015 A6M0001016 A6M0001017 A6M0001018 A6M0001019 A6M0001020 A6M0001021 A6M0001025 A6M0001026 A6M0001029 A6M0001022 A6M0001023 A6M0001024 A6M0001027 A6M0001028
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSA
Firm FEI
3007802293
Firm city
Maryland Heights
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register