Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Instruments Div. of Stryker Corporation: Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part number 11032). A software application used to help estimation of blood loss. Used with blood bearing canisters and accessories.

Recall number
Z-1402-2022
Recalling firm
Stryker Instruments Div. of Stryker Corporation
Product
Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part number 11032). A software application used to help estimation of blood loss. Used with blood bearing canisters and accessories.
Status
Terminated
Initiated
2020-01-22
Posted
not on file
Terminated
2024-01-29
Reason
Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in lower than expected results.
Root cause
Process control
Action
Customers were notified via email dated 1/22/20. We request the following actions from you: 1. Please confirm receipt of this communication by emailing myself and/or your Gauss representative. 2. Please identify any of these miscolored Canister Scanning Labels in your possession: " Find the the miscolored labels by locating any package of insert + scanning label with a date of manufacture prior to to 2019-12-24. See an example below: 3. Once identified, please: " Quarantine this product and indicate this has been completed by written email response to this communication Your Gauss representative will be in contact with you this week to help resolve this issue. Replacement product will be provided immediately. Thank you for your attention to this matter. Please feel free to reach out directly to me if you have any questions.
Quantity
16,700 units
Distribution
US Nationwide
Lot, serial or model codes
UDI-DI GTIN: 10859506006101, 10859506006125 All Canister Scanning Labels with a date of manufacture prior to to 2019-12-24
510(k) numbers
["K142801"]
PMA numbers
not on file
Product code
PBZ
Firm FEI
1811755
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register