Biomerieux Inc: Vitek 2 AST-P640 REF 418 579 Gram Positive Susceptibility Card, 20 cards per container, IVD. Intended for use with the Vitek 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae.

Recall number
Z-1402-2015
Recalling firm
Biomerieux Inc
Product
Vitek 2 AST-P640 REF 418 579 Gram Positive Susceptibility Card, 20 cards per container, IVD. Intended for use with the Vitek 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae.
Status
Terminated
Initiated
2015-02-25
Posted
2015-04-08
Terminated
2017-03-22
Reason
The outer carton label and package insert for the recalled product incorrectly lists Streptococcus pneumoniae as an organism for intended use.
Root cause
Employee error
Action
Biomerieux, Incs sent an Urgent Product Correction Notice on February 25, 2015, to all affected customers. The letter identified the product, the problem, and the actions to be taken by the customer. Customers were instructed to do the following: " Please confirm this letter has been distributed and reviewed by all appropriate personnel within your organization. " For the referenced test kits (VITEK 2 ASTP635, ASTP640 or ASTP641), review the claimed organisms in the Package Insert antibiotic table to confirm which organisms can be tested. " Refrain from using the referenced test kits to determine the susceptibility of Streptococcus pneumoniae. " Reference your Laboratory Standard Operating Procedures to identify card type used for specific organisms. " Please store this letter with your bioMrieux instrument documentation. Please complete the attached Acknowledgement Form and return it to bioMrieux, Inc. Customers with questions were instructed to contact their local BioMerieux representative. For questions regarding this recall call 314-731-8526.
Quantity
2053 cartons
Distribution
Worldwide Distribution to United Kingdom and Turkey. NO U.S. distribution.
Lot, serial or model codes
lot 740340920, expiration 14APR16; lot 740350820, expiration 22JUL16;
510(k) numbers
not on file
PMA numbers
not on file
Product code
LTW
Firm FEI
1950204
Firm city
Hazelwood
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register