Skip to the content

FDA Medical Device Recalls (openFDA)

Merit Medical Systems, Inc.: Prelude Short Sheath Introducer, Carolina Kidney and Endocrine - Declot tray, K12T-02693, Merit Medical Systems, Inc., South Jordan, Utah 84095. To provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. This device can also provide access to a native or synthetic graft used for hemodialysis. The side port of the sheath allows adequate flow to perform temporary hemodialysis.

Recall number
Z-1402-2009
Recalling firm
Merit Medical Systems, Inc.
Product
Prelude Short Sheath Introducer, Carolina Kidney and Endocrine - Declot tray, K12T-02693, Merit Medical Systems, Inc., South Jordan, Utah 84095. To provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. This device can also provide access to a native or synthetic graft used for hemodialysis. The side port of the sheath allows adequate flow to perform temporary hemodialysis.
Status
Terminated
Initiated
2009-04-14
Posted
2009-06-10
Terminated
2010-02-09
Reason
External tubing may separate during use, allowing blood loss.
Root cause
Process control
Action
All of Merit's US sales representatives and distributors were notified by Prelude Short Sheath Introducer - Product Recall letter on 4/14/2009. They were instructed to promptly contact their customers. Customers were instructed to isolate, remove from inventory, and return all affected inventory. A Product Retrieval Form was to be returned to Merit. Further information is available at 801-316-4998 or 801-208-4344.
Quantity
9 units
Distribution
Worldwide Distribution -- USA, including states of CA, CO, GA, IL, MO, OK, and OR and countries of Austria, Belgium, Canada, Denmark, France, Germany, Hong Kong, Japan, Latvia, Lithuania, Netherlands, New Zealand, Spain, Sweden, and UK.
Lot, serial or model codes
Lot number V686326.
510(k) numbers
["K082063"]
PMA numbers
not on file
Product code
DYB
Firm FEI
1721504
Firm city
South Jordan
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register