Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: ONCOR Impression Plus, Digital Linear Accelerator, Model No. 05857912 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.

Recall number
Z-1401-2018
Recalling firm
Siemens Medical Solutions USA, Inc
Product
ONCOR Impression Plus, Digital Linear Accelerator, Model No. 05857912 Product Usage: The intended use of the device is to deliver x-ray radiation for therapeutic treatment of cancer.
Status
Terminated
Initiated
2018-02-14
Posted
not on file
Terminated
2019-05-16
Reason
Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.
Root cause
Software design
Action
On February 16, 2018, the recalling firm issued a Customer Advisory Notice via USPS Certified Mail to inform customers about the update of the Control Console software version 9.2.400 of the Digital Linear Accelerator to version 9.2.502. Customers were informed of the issues that would be fixed through this software update. Consignees were asked to instruct affected personnel accordingly, place the safety advisory notice in the device's instructions for use, and have personnel maintain awareness until the fix is implemented.
Quantity
15
Distribution
Worldwide Distribution- US Nationwide in the states: AL, CA, FL, GA, IL, IN, KY, MA, MI, MO, MS, NJ, NY, OH, OK, PA, RI, SC, TN, TX, WI, WV, and WY; and PR. International distribution to Nassau.
Lot, serial or model codes
Serial no: 3855 4019 4067 3953 5148 3793 5098 5222 3801 5388 3948 3755 5033 4015 5095
510(k) numbers
["K121295"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register