Endocare, a Wholly Owned Subsidiary of HealthTronics, Inc.: V-Probe, Endocare Cryo V Probe, single unit model # CVA2400. The product is used in a minimally invasive, percutaneous approach to treating prostate cancer in low, moderate and high risk patients. Prostate Cryoablation uses extremely cold temperatures to destroy cancerous tissue.

Recall number
Z-1401-2010
Recalling firm
Endocare, a Wholly Owned Subsidiary of HealthTronics, Inc.
Product
V-Probe, Endocare Cryo V Probe, single unit model # CVA2400. The product is used in a minimally invasive, percutaneous approach to treating prostate cancer in low, moderate and high risk patients. Prostate Cryoablation uses extremely cold temperatures to destroy cancerous tissue.
Status
Terminated
Initiated
2008-12-05
Posted
2010-04-16
Terminated
2010-10-21
Reason
Endocare received a report of a patient death that occurred during a prostate cryoablation procedure. An autopsy attributed the cause of death to an air embolism. The safety notice for the cryoprobes was issued to refocus attention on the Indications and Safety Review included with the Physician Training Materials and to reinforce additional safety information.
Root cause
Other
Action
The recall was initiated with Endocare sending a notification letter to identified customers on December 5, 2008 and offering an optional replacement unit. The initial correspondence was sent out via an e-mail if the correspondence was returned undeliverable or an e-mail address was not available it was then sent Federal Express to those recipients. The notification letter explained the reason for recall and requested providers refocus their attention on the Indications & Safety Review information contained in the Endocare's Physician Training Materials. Physicians were instructed to contact the firm at 1-949-450-5412 to discuss any questions they may have about the information in the notification letter.
Quantity
22,741 units total for all products
Distribution
Nationwide.
Lot, serial or model codes
Lot Numbers 08-0539.
510(k) numbers
["K011074","K023757","K060279"]
PMA numbers
not on file
Product code
GEH
Firm FEI
3008262715
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register