Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp.: Integra Radionics HRAIM Intubation Head Ring Assembly Ref: HRAIM Head Rings serve as the general stereotactic treatment platform. Head Rings are used to provide a reference frame for instrumentation used for precise spatial localization and treatment of physiologic targets for stereotactic neurosurgical procedures such as craniotomies, biopsies, functional neurosurgery, and radiation therapy. Head Rings are delivered to the user non-sterile, and are reusable.

Recall number
Z-1400-2011
Recalling firm
Integra LifeSciences Corp.
Product
Integra Radionics HRAIM Intubation Head Ring Assembly Ref: HRAIM Head Rings serve as the general stereotactic treatment platform. Head Rings are used to provide a reference frame for instrumentation used for precise spatial localization and treatment of physiologic targets for stereotactic neurosurgical procedures such as craniotomies, biopsies, functional neurosurgery, and radiation therapy. Head Rings are delivered to the user non-sterile, and are reusable.
Status
Terminated
Initiated
2011-01-12
Posted
2011-02-23
Terminated
2011-04-04
Reason
Overall length of the intubation hoop in the HRAIM Intubation Head Ring Assembly is too long and will not allow a device to to attach
Root cause
Manufacturing material removal
Action
Integra sent a Medical Device Field Corrective Action letter dated January 12, 2011, to all affected customers via Federal Express. Integra Sales Specialists were provided with the replacement part and written instructions on how to replace the affected intubation hoops from customer inventory identified in the traceability report. Upon completion of the field correction each sales specialist was instructed to return the affected hoop along with a completed copy the Recall Acknowledgement and Return Form. If additional information is required contact me at (609) 936-2485 or james.tillman@integraJife.com.
Quantity
8 units
Distribution
Worldwide Distribution - USA, Canada, and Australia
Lot, serial or model codes
Lot/Serial Number(s): 0189463, 0189289, 0186642, 0184885
510(k) numbers
["K944463"]
PMA numbers
not on file
Product code
HAW
Firm FEI
3003418325
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register