Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions Diagnostics: Immulite 2000 and Immulite 2500 digital Fluidics PC Board, Part Number 450116-0002. Performs chemiluminescent immunoassays using serum, plasma, and/or urine for in vitro diagnostic testing.

Recall number
Z-1399-2009
Recalling firm
Siemens Medical Solutions Diagnostics
Product
Immulite 2000 and Immulite 2500 digital Fluidics PC Board, Part Number 450116-0002. Performs chemiluminescent immunoassays using serum, plasma, and/or urine for in vitro diagnostic testing.
Status
Terminated
Initiated
2009-01-05
Posted
2009-05-26
Terminated
2010-04-19
Reason
Design Flaw: The redesign of the Digital Fluidics PCB used in IMMULITE 2000 and IMMULITE 2500 is missing a jumper.
Root cause
Process design
Action
Siemens Healthcare Diagnostics Inc. issued an "Urgent: Device Recall" notice dated January 2009 mailed to all direct customers of the affected product. The letter described the problem, provided "work-around" instructions and also provided a Recall Response Form to be returned to the firm via fax. For further information, please contact your local Siemens Healthcare support representative or Siemens Healthcare Diagnostics Inc. directly at 1-973-927-2828.
Quantity
295 Digital Fludics PCB (88 IMMULITE 2000 and 2500 and 207 as spare parts).
Distribution
Worldwide Distribution to states including MA, NY, PA, MD, GA, FL, OH, IN, MI, IL, MO, LA, UT, AZ and CA and countries of Canada, Czech Republic, Germany, Poland, Hungary, Belgium, Russia, Republic of South Africa, Greece, Israel, Turkey, Spain, Portugal, Denmark, France, Great Britain, India, China, South Korea and Mexico.
Lot, serial or model codes
IMMULITE 2000: Serial Numbers M5080 to M5137; and IMMULITE 2500: Serial Numbers F0778 to F0818.
510(k) numbers
["K033234","K970227"]
PMA numbers
not on file
Product code
DGC
Firm FEI
1000122204
Firm city
Flanders
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register