FDA Medical Device Recalls (openFDA)
Becton Dickinson & Co.: BD Phoenix NMIC-306 Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells with dried biochemical substrates that use traditional qualitative microbiology methods such as fermentation, oxidation, degradation, and hydrolysis in combination with chromogenic and fluorogenic substrates along with carbon sources for bacterial identification and quantitation.
- Recall number
- Z-1398-2024
- Recalling firm
- Becton Dickinson & Co.
- Product
- BD Phoenix NMIC-306 Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) polystyrene panel (tray) containing control wells and substrate wells with dried biochemical substrates that use traditional qualitative microbiology methods such as fermentation, oxidation, degradation, and hydrolysis in combination with chromogenic and fluorogenic substrates along with carbon sources for bacterial identification and quantitation.
- Status
- Open, Classified
- Initiated
- 2024-02-29
- Posted
- 2024-03-29
- Terminated
- not on file
- Reason
- Products were found to have confirmed drug cross contamination prior to release may lead to incorrect, inadequate, or absent antibiotic treatment or delay in results
- Root cause
- Under Investigation by firm
- Action
- BD issued URGENT: Medical Device Product Correction Product Removal letter on 2/29/24 via courier service via tracked letter and/or e-mail. Letter states reason for recall, health risk and action to take: 1. Immediately inspect your inventory for the specific catalog and lot number listed above. Destroy affected product subject to the recall following your institution s process of destruction. 2. Ensure the contents of this notification are read and understood. 3. Share and post this customer letter within your facility network and forward to any customers to whom you may have distributed the product to ensure awareness. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements. BD will issue product replacements following receipt of the completed Customer Response Form. Contact: North American Regional Complaint Center Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Recall when prompted Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com Product Advisory, Recall Questions, Product Complaints, Technical Questions
- Quantity
- 11,800 eaches
- Distribution
- Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.
- Lot, serial or model codes
- UDI-DI: (10) 3318700 (17)241107 (30)0025 (01) 30382904492923 Lot Number: 3318700 Exp. Date: 07-NOV-2024
- 510(k) numbers
- ["K132674"]
- PMA numbers
- not on file
- Product code
- LON
- Firm FEI
- 1119779
- Firm city
- Sparks
- Firm state
- MD
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28