FDA Medical Device Recalls (openFDA)
Orthovita, Inc., dBA Stryker Orthobiologics.: Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Beveled Needle, 6", Part Number 2110-0524 Product Usage: Facilitate placement of bone cement in weakened or diminished bone
- Recall number
- Z-1398-2013
- Recalling firm
- Orthovita, Inc., dBA Stryker Orthobiologics.
- Product
- Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Beveled Needle, 6", Part Number 2110-0524 Product Usage: Facilitate placement of bone cement in weakened or diminished bone
- Status
- Terminated
- Initiated
- 2012-11-10
- Posted
- 2013-05-24
- Terminated
- 2014-10-29
- Reason
- There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use.
- Root cause
- Labeling design
- Action
- Stryker sent an Urgent Medical Device Correction letter dated November 10, 2012 to all affected customers. The letter identified the affected product, problem issue, potential hazards and actions to be taken. Customers were instructed to immediately review this notification and the attached IFU. Customers were asked to complete and sign the enclosed Business Reply Form and fax a copy to 1-484-323-8803. For questions contact Stryker Orthobiologics.
- Quantity
- 12
- Distribution
- USA Nationwide Distribution in the states of AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, ND, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI.
- Lot, serial or model codes
- Manufacturer Part Number 2110-0524, Lot Number E909009R and E909009R
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OAR
- Firm FEI
- 1000307073
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as