Skip to the content

FDA Medical Device Recalls (openFDA)

B Braun Medical Inc: Introcan Safety 3 Closed IV Catheter 22 Ga. x 1 in., PUR, Winged. Passive safety IV catheter.

Recall number
Z-1397-2025
Recalling firm
B Braun Medical Inc
Product
Introcan Safety 3 Closed IV Catheter 22 Ga. x 1 in., PUR, Winged. Passive safety IV catheter.
Status
Open, Classified
Initiated
2025-02-17
Posted
2025-03-19
Terminated
not on file
Reason
Increased risk for valve housing detachment causing leakage.
Root cause
Process control
Action
An URGENT MEDICAL DEVICE RECALL NOTIFICATION dated 2/13/25 was mailed to consignees notifying them of this product removal. Consignees are instructed to review the recall notification and to forward the notice along to those in their organization and to those to which product was further distributed. Consignees are to quarantine any affected devices in their inventory and coordinate product return by completion and return of the provided Product Recall Acknowledgement Form. Consignees with any questions are to contact B Braun at 844-903-6417.
Quantity
28,200 units
Distribution
Domestic: CA, CA, FL, GA, IL, IL, IN, MD, MN, MO, NC, NM, NY, PA, TN, TX, VA, and WA; Foreign: Canada.
Lot, serial or model codes
Model No 4251128-02; UDI-DI 04046964014017; Lot 24F09G8952.
510(k) numbers
["K182870","K213085","K111236"]
PMA numbers
not on file
Product code
FOZ
Firm FEI
2521402
Firm city
Bethlehem
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register