Skip to the content

FDA Medical Device Recalls (openFDA)

Calyxo, Inc.: CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.

Recall number
Z-1396-2025
Recalling firm
Calyxo, Inc.
Product
CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.
Status
Open, Classified
Initiated
2025-02-19
Posted
2025-03-21
Terminated
not on file
Reason
Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.
Root cause
Device Design
Action
On 2/19/2025, correction notices were mailed to distributors and healthcare providers who were asked to do the following: 1) Notify all urologists utilizing the affected aspiration system in your institution that additional instructions are required. Please provide each of them with a copy of the additional instructions. 2) As another safety step, a firm representative will be contacting all users from your institution directly to offer training on these additional instructions. If you prefer to coordinate the training, please specifically state that preference in the acknowledgement. 3) Please acknowledge receipt of the notice, additional instructions, and updated training instructions MA00100.A by emailing qualitycontrol@calyxoinc.com, or calling 833-214-3354 and providing your name, title, institution and address.
Quantity
11,246
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV.
Lot, serial or model codes
CVC127020-1 UDI-DI: 00860005357710, L00018 Rev C
510(k) numbers
["K233472"]
PMA numbers
not on file
Product code
FED
Firm FEI
3014683069
Firm city
Pleasanton
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register