Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho-Clinical Diagnostics, Inc.: VITROS Immunodiagnostic Products Anti-SARSCoV- 2 Total N Antibody Calibrators programmed onto the Assay Data Disks (ADD) Data Release Versions (DRV): DRVs 6213 through 6218 supporting VITROS Calibrators. Product Code: 6199976

Recall number
Z-1396-2023
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Immunodiagnostic Products Anti-SARSCoV- 2 Total N Antibody Calibrators programmed onto the Assay Data Disks (ADD) Data Release Versions (DRV): DRVs 6213 through 6218 supporting VITROS Calibrators. Product Code: 6199976
Status
Open, Classified
Initiated
2023-03-09
Posted
2023-04-18
Terminated
not on file
Reason
Incorrect Number of Calibrator Levels (2) Set for the VITROS Immunodiagnostics Products Anti-SARS-CoV-2 Total N Antibody Calibrators in Assay Data Disks (ADD), resulting in delayed patients results
Root cause
Employee error
Action
Ortho Clinical Diagnostics issued Important Product Correction Notification letter on 3/9/23. Letter states reason for recall, health risk and action to take: To calibrate all lots currently within expiring date of Anti-SARS-CoV-2 Total N, load ADD 6219 or higher on your VITROS System. " Complete the enclosed Confirmation of Receipt form no later than March 17, 2023. " Please forward this notification and ADD if the affected product was distributed outside of your facility. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Ortho Care" Technical Solutions Center. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
Quantity
256 units
Distribution
US Nationwide distribution in the states of AL, AZ, CA, GA, NY.
Lot, serial or model codes
UDI: 10758750034581 Lot # /Expiration Date: 0230 22-Feb-2023; 0240 22-Mar-2023; 0250 26-Apr-2023.
510(k) numbers
not on file
PMA numbers
not on file
Product code
QKO
Firm FEI
2250051
Firm city
Raritan
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register