FDA Medical Device Recalls (openFDA)
Mckesson Medical Imaging: McKesson Radiology 12.2 - Picture Archive Communication System (PACS)
- Recall number
- Z-1396-2017
- Recalling firm
- Mckesson Medical Imaging
- Product
- McKesson Radiology 12.2 - Picture Archive Communication System (PACS)
- Status
- Terminated
- Initiated
- 2016-12-02
- Posted
- not on file
- Terminated
- 2017-11-24
- Reason
- Issue for customers that use an EMR login or legacy web URL login or legacy web URL login for McKesson Radiology PACS that may result in missing images in a newly imported study, and/or study imports that remain in an "in-progress" status.
- Root cause
- Software design
- Action
- McKesson sent an Urgent Field Safety Notice on December 2, 2016 to customers with potentially affected software versions. The notice indicated that they were to contact McKesson to install a software update to correct the issue, and that they read and understood the Field Safety Notice.
- Quantity
- 17
- Distribution
- AK, CA, FL, HI, KY, MD, MI, MS, NH, OH, PA, TN, TX, VT
- Lot, serial or model codes
- Model 12.2
- 510(k) numbers
- ["K140909"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 3003757840
- Firm city
- Richmond
- Firm state
- not on file
- Firm country
- Canada
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28