Skip to the content

FDA Medical Device Recalls (openFDA)

MED-EL Elektromedizinische Gereate, Gmbh: SONNET Mini Battery Pack CableProduct Usage: The SONNET Mini Battery Pack Cable is part of the SONNET Mini Battery Pack, which is an external battery pack used to power one MED-EL SONNET control unit. It can be worn on the upper body or the hip using different SONNET Mini Battery Pack Cable lengths to connect the SONNET Mini Battery Pack to one processor unit.

Recall number
Z-1395-2017
Recalling firm
MED-EL Elektromedizinische Gereate, Gmbh
Product
SONNET Mini Battery Pack CableProduct Usage: The SONNET Mini Battery Pack Cable is part of the SONNET Mini Battery Pack, which is an external battery pack used to power one MED-EL SONNET control unit. It can be worn on the upper body or the hip using different SONNET Mini Battery Pack Cable lengths to connect the SONNET Mini Battery Pack to one processor unit.
Status
Terminated
Initiated
2016-10-25
Posted
2017-03-02
Terminated
2017-04-04
Reason
Unit Defect: Firm inspection revealed units did not have the compartment lock firmly welded to the housing.
Root cause
Device Design
Action
MED-EL is notifying all consignees to whom affected products have been directly shipped by Headquarters (based on first delivery shipment records) on 10/25/2016.
Quantity
538 units
Distribution
Worldwide Distribution - US Nationwide in the states of CA, NC and TX; and the countries of Belgium, Belarus, Canada, France, Germany, Hong Kong, Israel, Netherlands, Philippines, Poland, Portugal, Turkey, United Kingdom, and Uruguay.
Lot, serial or model codes
Part No. 32291, Kit 33372, Kit UDI - Primary ID: 09008737333720 (anthracite 50); Part No. 32292, Kit 33374, Kit UDI - Primary ID: 09008737333744 (ebony 50); Part No. 32293, Kit 33375, Kit UDI - Primary ID: 09008737333751 (black 50); Part No. 32294, Kit 33376, Kit UDI - Primary ID: 09008737333768 (white 50); Part No. 32295, Kit 33377, Kit UDI - Primary ID: 09008737333775 (beige 50) and Part No. 32296, Kit 33378, Kit UDI - Primary ID: 09008737333782 (grey 50).
510(k) numbers
not on file
PMA numbers
["P000025S087"]
Product code
ISM
Firm FEI
3000249315
Firm city
Innsbruck
Firm state
not on file
Firm country
Austria

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28