Skip to the content

FDA Medical Device Recalls (openFDA)

C.R. Bard, Inc.: BARD ArcticGel Pads Product Usage: The Arctic Sun Temperature Management System is intended for monitoring and controlling patient temperature. ArcticGel Pads are only for use with an Arctic Sun Temperature Management System Control Module.

Recall number
Z-1395-2016
Recalling firm
C.R. Bard, Inc.
Product
BARD ArcticGel Pads Product Usage: The Arctic Sun Temperature Management System is intended for monitoring and controlling patient temperature. ArcticGel Pads are only for use with an Arctic Sun Temperature Management System Control Module.
Status
Terminated
Initiated
2016-03-04
Posted
not on file
Terminated
2017-12-07
Reason
Complaints were received of restricted/inaccurate flow rate.
Root cause
Process control
Action
Bard Medical sent an Urgent Medical Product Recall letter dated March 4, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to not use any of the affected products and to return any unused product Bard Medical Division via the provided Customer Recall and Effectiveness Form. BMD will be providing a replacement for the returned product. If the product was further distributed, the consignee was instructed to identify their customers and notify them of the product recall.
Quantity
21,412 units
Distribution
Worldwide Distribution - Domestic: US Nationwide; Foreign: Argentina, Australia, Austria, Belarus, Belgium, Bosnia, Brazil Canada, Chile, China, Colombia, Croatia, Cyprus, Denmark, Egypt, Finland, France, Germany, Greece, Oman Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, , South Korea, Kuwait, Macedonia, Malaysia, Mexico, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, The Netherlands, Turkey, United Arab Emirates, Ukraine, United Kingdom, Uruguay and Vietnam.
Lot, serial or model codes
All lots with Catalog numbers: 31700 (Universal Pad), 3170004 (Universal Pad 4 pk), 31703 (XSmall pad kit), 3170302 (XSmall pad kit 2 pk), 31705 (Small pad kit), 3170502 (Small pad kit 2 pk), 31707 (Medium pad kit), 3170702 (Medium pad kit 2 pk), 31709 (Large pad kit) and 3170902 (Large pad kit 2 pk), that are within their current expiration dates.
510(k) numbers
["K142702"]
PMA numbers
not on file
Product code
DWJ
Firm FEI
1018233
Firm city
Covington
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register