Skip to the content

FDA Medical Device Recalls (openFDA)

Bio-Rad Laboratories Inc: Bio-Rad Variant II Hemoglobin Testing System with CDM v 4.0, model number 270-2000, Manufactured by Bio-Rad Laboratories, Hercules, CA Measurement of percent hemoglobin Alc is effective in monitoring long-term glucose control in individuals with diabetes mellitus.

Recall number
Z-1395-2009
Recalling firm
Bio-Rad Laboratories Inc
Product
Bio-Rad Variant II Hemoglobin Testing System with CDM v 4.0, model number 270-2000, Manufactured by Bio-Rad Laboratories, Hercules, CA Measurement of percent hemoglobin Alc is effective in monitoring long-term glucose control in individuals with diabetes mellitus.
Status
Terminated
Initiated
2008-12-09
Posted
2009-05-19
Terminated
2010-11-12
Reason
The wrong barcode number and tube position number may be assigned to samples leading to discrepant sample results.
Root cause
Pending
Action
Recall initiated on December 10, 2008. Notification conducted by phone call and fax for domestic consignees. Each subsidiary will receive the subsidiary Urgent: Medical Device Correction notice, customer medical device correction response form and number of customers distributed to by e-mail. Each subsidiary must contact customers and provide the customer notice and response form as needed. The notice stated that Bio-Rad Laboratories would release a follow up communication providing a permanent solution upon completion of an investigation. In the interim, it was recommended that customers carefully review the Summary Report for duplicate barcode IDs or or tube position numbers within the same rack. If there are any questions or assistance needed, the local regional Bio-Rad office should be contacted.
Quantity
42 units
Distribution
Worldwide Distribution -- USA, Argentina, Australia, Brazil, Canada, China, Czech Republic, France, Germany, Hungary, India, Italy, Mexico, New Zealand, Singapore, South Korea, Sweden, Thailand, and the UK.
Lot, serial or model codes
CDM version 4.0 software
510(k) numbers
["K063643"]
PMA numbers
not on file
Product code
LCP
Firm FEI
2915274
Firm city
Hercules
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register