Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Renal Div: Baxter Arena Hemodialysis Systems, product codes ARENADPX, ARENASP, ARENADPP, ARENASPX; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085

Recall number
Z-1395-05
Recalling firm
Baxter Healthcare Renal Div
Product
Baxter Arena Hemodialysis Systems, product codes ARENADPX, ARENASP, ARENADPP, ARENASPX; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085
Status
Terminated
Initiated
2005-06-14
Posted
2005-08-24
Terminated
2008-02-26
Reason
Operator's Manuals addenda are being issued to provide correct information concerning the correct syringe size to use for heparin infusion, the use of chemical disinfectants in the System 1000, TMP Advise, and Bicarbonate Value calculations.
Root cause
Other
Action
Baxter sent Important Product Information letters dated 6/14/05 to all System 1000 and Arena Hemodialysis customers, to the attention of the Hemodialysis Administrator, via first class mail. Enclosed with the letter were copies of addenda to the Operator''s Manual for both the System 1000 and Arena instruments. The letter informed the accounts of the following problems: a) Improper Syringe Size - it is important to use the correct syringe size for heparin infusion to avoid improper infusion of heparin. The warning 'Ensure the instrument is properly configured for the type and size of heparin pump syringe being installed.' was added to the manual under the Priming the Dialyzer and Blood Lines section of Operation. b) Disinfection Infusion Monitor (System 1000 only) - Procedural steps have been added to the System 1000 manual to confirm the level in the disinfectant bottle prior to and after chemical disinfection to ensure that disinfection fluid has been utilized during the disinfection cycle. c) TMP Advise - the 'TMP Alarm in Dialyzer Mode' section in both the System 1000 and Arena manuals has been removed. d) Bicarbonate Value - the Dialysate Preparation section of the manual was revised to clarify the bicarbonate display of the System 1000 and Arena instruments. Any questions were directed to Baxter Global Technical Services at 1-800-553-6898.
Quantity
623 units
Distribution
Nationwide and internationally to Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, India, Indonesia, Korea, Mexico, Peru, Philippines, Singapore, Spain, Switzerland, Taiwan, Thailand and Turkey.
Lot, serial or model codes
all serial numbers
510(k) numbers
["K030099"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3000210122
Firm city
Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register