Skip to the content

FDA Medical Device Recalls (openFDA)

Alcon Research, Ltd.: VERION Reference Unit (Vision Planner) Product Usage: The VERION Image Guided system is an advanced surgical planning, imaging and guidance technology designed to provide greater accuracy and efficiency during cataract surgery. It allows eye surgeons to capture a high-resolution, diagnostic reference image of the patients eye pre-operatively, quickly determine an optimized surgical plan that enables surgeons to see all inclusions and alignment in real-time.

Recall number
Z-1394-2016
Recalling firm
Alcon Research, Ltd.
Product
VERION Reference Unit (Vision Planner) Product Usage: The VERION Image Guided system is an advanced surgical planning, imaging and guidance technology designed to provide greater accuracy and efficiency during cataract surgery. It allows eye surgeons to capture a high-resolution, diagnostic reference image of the patients eye pre-operatively, quickly determine an optimized surgical plan that enables surgeons to see all inclusions and alignment in real-time.
Status
Terminated
Initiated
2016-03-01
Posted
2016-04-14
Terminated
2017-04-13
Reason
Alcon is conducting a voluntary medical device correction of all VERION Reference Units (Vision Planner) that are shared with the Alcon LenSx Laser System after receiving reports concerning the inclusion of unplanned arcuates on printed, saved (.pdf) or exported surgical plans.
Root cause
Software change control
Action
Alcon sent an Urgent Medical Device Correction letter dated February 29, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that an Alcon Technical Services Engineers will contact customers, schedule Service Calls, and perform onsite corrections. For questions call 1-800-862-5266.
Quantity
97 units
Distribution
Worldwide Distribution - US Nationwide throughout the US and the countries of: Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Czech Republic, Chile, China, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Greece, Guatlemala, Hungary, India, Indoesia, Isreal, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Nicaragua, Norway, Panama, Philippines, Portugal, Romania, Singapore, Slovakia, South Africa, Spain, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, UK, Ukrain, United Arab Emirates
Lot, serial or model codes
Serial Numbers: 1017, 1119, 1123, 1125, 1126, 1128, 1133, 1135, 1136, 1137, 1138, 1139, 1142, 1147, 1150, 1153, 1155, 1161, 1165, 1168, 1172, 1178, 1181, 1182, 1183, 1188, 1190, 1191, 1193, 1196, 1197, 1198, 1199, 1201, 1203, 1206, 1207, 1220, 1226, 1228, 1230, 1232, 1250, 1256, 1258, 1259, 1421, 1433, 1435, 1439, 1451, 1452, 1506, 1508, 1509, 1523, 1536, 1548, 1559, 1562, 1565, 1568, 1581, 1597, 1606, 1628, 1632, 1649, 1661, 1663, 1668, 1671, 1694, 1696, 1705, 1716, 1726, 1729, 1798, 1816, 1818, 1820, 1824, 1829, 1921, 1922, 1934, 1937, 1938, 1940, 1942, 1954, 1957, 2057, 2069, 2086, 2092
510(k) numbers
not on file
PMA numbers
not on file
Product code
NFG
Firm FEI
1610287
Firm city
Fort Worth
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.