Microvention, Inc.: MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft 3D, Sterile, Rx, REF numbers 100102HS3D-V, 100103HS3D-V, 100152HS3D-V, 100153HS3D-V, 100154HS3D-V, 100202HS3D-V, 100203HS3D-V, 100204HS3D-V, 100206HS3D-V, 100208HS3D-V, 100254HS3D-V, 100256HS3D-V, 100304HS3D-V, 100306HS3D-V, 100310HS3D-V, 100355HS3D-V, 100358HS3D-V, 100406HS3D-V, 100408HS3D-V, 100412HS3D-V, 100415HS3D-V, 100510HS3D-V, 100515HS3D-V; MV-00102HTDA, MV-00103HTDA, MV-00204HTDA,MV-00304HTDA, MV-00306HTDA, MV-00406HTDA, MV-00408HTDA, MV-00510HTDA, MV-01502HTDA, MV-01503HTDA, MV-01504HTDA; MV-02504HTDA, MV-02506HTDA, MV-03505HTDA, and MV-03508HTDA. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.

Recall number
Z-1393-2020
Recalling firm
Microvention, Inc.
Product
MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft 3D, Sterile, Rx, REF numbers 100102HS3D-V, 100103HS3D-V, 100152HS3D-V, 100153HS3D-V, 100154HS3D-V, 100202HS3D-V, 100203HS3D-V, 100204HS3D-V, 100206HS3D-V, 100208HS3D-V, 100254HS3D-V, 100256HS3D-V, 100304HS3D-V, 100306HS3D-V, 100310HS3D-V, 100355HS3D-V, 100358HS3D-V, 100406HS3D-V, 100408HS3D-V, 100412HS3D-V, 100415HS3D-V, 100510HS3D-V, 100515HS3D-V; MV-00102HTDA, MV-00103HTDA, MV-00204HTDA,MV-00304HTDA, MV-00306HTDA, MV-00406HTDA, MV-00408HTDA, MV-00510HTDA, MV-01502HTDA, MV-01503HTDA, MV-01504HTDA; MV-02504HTDA, MV-02506HTDA, MV-03505HTDA, and MV-03508HTDA. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Status
Terminated
Initiated
2019-11-22
Posted
not on file
Terminated
2022-01-26
Reason
The devices may be missing the implant coil.
Root cause
Process design
Action
The recalling firm issued letters beginning 11/22/2019 dated 11/26/2019 via email using different flags according to the region the foreign consignee was located, i.e. 'URGENT: FIELD SAFETY CORRECTION ACTION" (for EMEA), "URGENT: MEDICAL DEVICE RECALL" (for Japan), or "FIELD SAFETY ADVISORY NOTICE" (for Rest of World).
Quantity
1,725 units
Distribution
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
Lot, serial or model codes
Catalog/lot numbers:  100102HS3D-V - 1903045WT, 1903065WT, 1903205WT, 1904015WT, 1904035WT, 1904245WT, 1904255WT, 1904265WT, 1904295WT, 1905025WT, 1905065WT, 1905085WT, 1905155WT, 1905175WT, and 1905205WT;  100103HS3D-V - 1903115TT, 1903015WT, 1903085WT, 1903135WT, 1904015WT, 1904035WT, 1904055WT, 1904105WT, 1904265WT, 1904295WT, 1905035WT, 1905105PT, 1905135WT, 1905155WT, and 1905225WT;  100152HS3D-V - 1903065NT, 1903045WT, 1903085WT, 1903205WT, 1903225WT, 1904015WT, 1904105WT, 1904175WT, 1904295WT, 1905025WT, 1905105WT, 1905135WT, 1905155WT, and 1905175WT;  100153HS3D-V - 1903015WT, 1903045WT, 1903065WT, 1903155WT, 1904015WT, 1904035WT, 1904265WT, 1904295WT, 1905135WT, and 1905205WT;  100154HS3D-V - 1903045WT, 1903085WT, 1903135WT, 1903185WT, 1903225WT, 1904015WT, 1904035WT, 1904085WT, 1904105WT, 1904155WT, 1904175WT, 1904225WT, 1905025WT, 1905135WT, 1905155PT, 1905155WT, 1905175WT, and 1905205WT;  100202HS3D-V - 1903015WT, 1903065WT, 1903185WT, 1903205WT, 1905105WT, and 1905205WT;  100203HS3D-V - 1904015WT, 1904035WT, 1904265WT, 1904295WT, 1905065WT, 1905085WT, 1905155WT, and 1905225PT;  100204HS3D-V - 1903015WT, 1903045WT, 1903065WT, 1903085WT, 1904035WT, 1904055WT, 1904085WT, 1904245WT, 1904255WT, 1904295WT, 1905035WT, 1905065WT, and 1905085WT;  100206HS3D-V - 1904175WT, 1904225WT, 1904245WT, 1905105WT, 1905155PT, and 1905155WT;  100208HS3D-V - 1903085WT, 1903155WT, 1904035WT, 1905155PT, and 1905155WT;  100254HS3D-V - 1903015WT, 1903065WT, 1904015WT, 1904035WT, 1904055WT, 1904155WT, 1904225WT, 1904245WT, 1905035PT, 1905065WT;  100256HS3D-V - 1903015WT, 1903045WT, 1903065WT, 1903085WT, 1904015WC, 1904015WT, 1904035WT, 1904175WT, 1904245WT, 1904255WT, 1904265WT, 1904295WT, 1905035WH, 1905035WT, 1905155PT, and 1905225WT;  100304HS3D-V - 1903065WT, 1903115WT, 1903155WT, 1903205WT, 1904015WT, 1904055WT, 1904255WT, 1905105WT, 1905175WT, and 1905205WT;  100306HS3D-V - 1903085NT, 1903015WT, 1903085WT, 1903115WT, 1903135WT, 1903155WT, 1904015WT, 1904035WT, 1904055WT, 1904085WT, 1904225PT, 1904225WT, 1904245WT, 1904255WT, 1904265WT, 1905135WT, 1905175WT, and 1905205WT;  100310HS3D-V - 1903085WT and 1904035WT;  100355HS3D-V - 1903045WT, 1903085WT, 1904015WT, 1904225PT, 1904225WT, and 1905175WT;  100358HS3D-V - 1903065WT, 1904255WT, 1904265WT, 1904295WT, 1905035WT, 1905065WT, 1905085WT, 1905155WT, and 1905175WT;  100406HS3D-V - 1903015WT, 1903115WT, 1903205WT, 1904015WT, and 1905205WT;  100408HS3D-V - 1905205NT,1903085WT, 1903205WT, 1904015WT, 1904035WT, 1904055WT, 1904175WT, 1904225PT, 1904225WT, 1904255WT, 1904265WT, 1904295WT, 1905025WT, 1905105WT, 1905155WT, and 1905175WT;  100412HS3D-V - 1903135WT, 1904035WT, 1904175WT, and 1905105WT;  100415HS3D-V - 1903135WT, 1904035WT, 1905105WT, and 1905155WT;  100510HS3D-V - 1903065WT, 1904015WT, 1904035WT, 1904245WT, 1905105WT, and 1905155WT;  100515HS3D-V - 1903085NT, 1903015WT, 1903045WT, 1903085WT, 1903185WT, 1903205WT, 1904015WT, 1904035WT, 1904175WT, 1904265WT, 1904295WT, 1905025WT, 1905035WT, 1905085WT, 1905175WT;  MV-00102HTDA - 1903015YT, 1903045YT, 1903155XT, 1904105YT, and 1904155XT;  MV-00103HTDA - 1903045YT, 1903185YT, 1905065YT, and 1905085YT;  MV-00204HTDA - 1904085YT and 1905065YT;  MV-00304HTDA - 1905085YT;  MV-00306HTDA - 1903015YT, 1903155XT, and 1903155YT;  MV-00406HTDA - 1903015YT, 1904055YT, 1905065YT, and 1905085YT;  MV-00408HTDA - 1903155YT, 1904015XT, 1904085YT, 1905035XT, 1905065YT, and 1905085YT;  MV-00510HTDA - 1904085YT and 1905085YT;  MV-01502HTDA - 1903015YT, 1904085YT, 1904105YT, 1905085YT, and  1905105YT;  MV-01503HTDA - 1904055YT, 1904085YT, 1905085YT, 1905105XT, and 1905135YT;  MV-01504HTDA - 1904085YT, 1904105YT, 1905085YT, 1905105XT, 1905105YT, 1905135YT, and 1905155YT;  MV-02504HTDA - 1903155YT, 1904055YT, and 1905085YT;  MV-02506HTDA - 1903155YT, 1905085YT, and 1905105YT; and   MV-03505HTDA - 1903015YT, 1903155YT, 1904085YT, 1905085YT, and 1905105YT; and  MV-03508HTDA - 1904055YT, 1904085YT, 1905085YT, and 1905105XT.
510(k) numbers
["K050954","K091641","K132317","K132952","K153594","K161452"]
PMA numbers
not on file
Product code
HCG
Firm FEI
3003523199
Firm city
Aliso Viejo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register