Skip to the content

FDA Medical Device Recalls (openFDA)

Advanced Sterilization Products: Tyvek Self-Seal Pouch with STERRAD Chemical Indicator 12332. Intended to be used to contain medical devices that are being terminally sterilized in STERRAD Sterilization Systems.

Recall number
Z-1393-2011
Recalling firm
Advanced Sterilization Products
Product
Tyvek Self-Seal Pouch with STERRAD Chemical Indicator 12332. Intended to be used to contain medical devices that are being terminally sterilized in STERRAD Sterilization Systems.
Status
Terminated
Initiated
2010-06-23
Posted
2011-02-22
Terminated
2012-03-14
Reason
The recall was initiated after Advanced Sterilization Products (ASP) discovered a non-conforming lot. The misaligned perforations may have caused the seal to be incomplete on some of the pouches, creating a potential breach for pouches being used to store sterilized equipment on shelves.
Root cause
Process control
Action
Advanced Sterilization Products (ASP) released Urgent: Product Recall notifications to its consignees identifying the affected product and explaining the reason for recall. The letter also discussed the impact of using this affected product and return instructions. The firm asked customers to examine their inventory, discontinue use of Tyvek Self-Seal Pouch with STERRAD Chemical Indicator 12332 and to return all unused affected products to Stericycle. Customers are to complete the Business Reply Card and return it to Stericycle.
Quantity
470
Distribution
Worldwide Distribution -- USA, Argentina, Australia, Canada, Chile, Czech Republic, Hong Kong, India, Japan, Panama, Peru and Puerto Rico.
Lot, serial or model codes
Lot Number: 2020
510(k) numbers
["K951295"]
PMA numbers
not on file
Product code
KCT
Firm FEI
3003662624
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register