Skip to the content

FDA Medical Device Recalls (openFDA)

Microvention, Inc.: MicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, Helical, Sterile, Rx, REF numbers 100101H2HS-V; 100102H2HS-V, 100103H2HS-V, 100104H2HS-V, 100151H2HS-V; 100152H2HS-V; 100153H2HS-V; 100154H2HS-V; 100201H2HS-V; 100202H2HS-V; 100203H2HS-V; 100204H2HS-V; 100206H2HS-V, 100208H2HS-V, 100254H2HS-V, 100256H2HS-V, 100304H2HS-V; 100306H2HS-V; 100308H2HS-V; 100310H2HS-V; 100404H2HS-V; 100406H2HS-V, 100408H2HS-V, 100410H2HS-V, 100506H2HS-V, 100508H2HS-V; 100510H2HS-V, 100515H2HS-V, 100520H2HS-V, 100620H2HS-V, and MV-00208HHSA - Product Usage: The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.

Recall number
Z-1392-2020
Recalling firm
Microvention, Inc.
Product
MicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, Helical, Sterile, Rx, REF numbers 100101H2HS-V; 100102H2HS-V, 100103H2HS-V, 100104H2HS-V, 100151H2HS-V; 100152H2HS-V; 100153H2HS-V; 100154H2HS-V; 100201H2HS-V; 100202H2HS-V; 100203H2HS-V; 100204H2HS-V; 100206H2HS-V, 100208H2HS-V, 100254H2HS-V, 100256H2HS-V, 100304H2HS-V; 100306H2HS-V; 100308H2HS-V; 100310H2HS-V; 100404H2HS-V; 100406H2HS-V, 100408H2HS-V, 100410H2HS-V, 100506H2HS-V, 100508H2HS-V; 100510H2HS-V, 100515H2HS-V, 100520H2HS-V, 100620H2HS-V, and MV-00208HHSA - Product Usage: The HES is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.
Status
Terminated
Initiated
2019-11-22
Posted
not on file
Terminated
2022-01-26
Reason
The devices may be missing the implant coil.
Root cause
Process design
Action
The recalling firm issued letters beginning 11/22/2019 dated 11/26/2019 via email using different flags according to the region the foreign consignee was located, i.e. 'URGENT: FIELD SAFETY CORRECTION ACTION" (for EMEA), "URGENT: MEDICAL DEVICE RECALL" (for Japan), or "FIELD SAFETY ADVISORY NOTICE" (for Rest of World).
Quantity
2,762 units
Distribution
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
Lot, serial or model codes
Catalog/lot numbers:  100101H2HS-V - 1903135UR, 1903185UR, 1903225UR, and 1904035UR;  100102H2HS - V -1903135UR, 1903155UR, 1903155ZR, 1903185UR, 1903205UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR,1904155UR,  1904175UR, 1904175ZR, and 1904245UR;  100103H2HS-V - 1903015UR, 1903045UR, 1903065UR, 1903085UR, 1903135UR, 1903205UR 1904055UR, 1904085UR, 1904155UR, 1905085UR,  1905105UR, 1905135UR, and 1905155UR;  100104H2HS-V - 1903015UR, 1903135UR, 1903205UR, 1904085UR, 1904105UR, and 1904155UR;  100151H2HS-V - 1903085UR, 1903185UR, and 1904015UR;  100152H2HS-V - 1903135UR, 1903155UR, 1903205NR, 1903205UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR, 1904155UR, 1904175UR,  1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR, 1905035UR, 1905065UR, 1905085UR, 1903045WR, 1903185WR, 1903205WR, 1904105WR, 1904175WR, 1904225WR, 1905065WR, and 1905175WR;  100153H2HS-V - 1903015UR, 1903065UR, 1903085UR, 1903185UR, 1904015UR, 1904015ZR, 1904105UR, 1904155UR, 1904245UR, 1904265UR, and 1904295UR;  100154H2HS-V - 1903135UR, 1903155UR, 1903205UR, 1903225UR, 1904015UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR, 1904155UR,  1904175UR, 1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR,  1905035UR, 1905065NR, 1905065UR, 1905155NR, 1903185WR, 1903205WR, 1904105WR, 1904225WR, and 1905065WR;  100201H2HS-V - 1905105PR;  100202H2HS-V - 1903155UR, 1904105UR, 1903185WR, 1903205WR, 1904175WR, 1905065WR, 1905135WR, and 1905155WR;  100203H2HS-V - 1903015UR, 1904015ZR, 1904105UR, 1904155UR, 1904175UR, 1904225UR, 1905025UR, 1905035NR, 1905035UR, 1905065UR, 1905085UR, 1905135NR, 1905155NR, 1903185WR, and 1905035WR;  100204H2HS-V - 1903015UR, 1903065UR, 1903085UR, 1904035UR, 1904155UR, 1904175UR, 1904175ZR, 1904225UR, 1904245UR, 1904245ZR, 1904265UR, 1904295UR, 1905025UR, 1905035NR, 1905035UR, 1905065UR, 1905085UR, 1905135NR, 1905155NR, 1904105WR, 1904175WR, 1905035WR, 1905065WR, 1905085WR, 1905135WR, and 1905155WR;   100206H2HS-V - 1903045UR, 1903065UR, 1903205UR, 1903225UR, 1904035UR, 1904055UR, 1904085UR, 1904105UR, 1904155UR, 1904175UR,  1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR, 1905035UR, 1905065NR, 1905065UR, 1905085UR, 1905105UR, 1905135UR, 1903185WR, 1904225WR, 1905065WR, 1905105PR, and 1905135WR;  100208H2HS-V - 1903015UR, 1903045UR, 1903085UR, 1903205UR, 1903225UR, 1904015UR 1904035UR, 1904175ZR, 1904225UR, 1904245UR, 1904265UR, 1904295UR, 1905025UR, 1905035UR, 1905065UR, 1905085UR, 1905105UR, 1905135UR, and 1905085WR;  100254H2HS-V - 1904245NR, 1903185WR, and 1905065WR;  100256H2HS-V - 1905065WR;  100304H2HS-V - 1903065WR, 1903205WR, and 1904225WR;  100306H2HS-V - 1903065UR, 1903085UR, 1904155UR, 1904175UR, 1904225UR, 1905135NR, 1905035WR, 1905065WR, 1905105PR, 1905135WR, and 1905155WR;  100308H2HS-V - 1904015UR, 1904155UR, 1904175UR, 1904225UR, 1904265UR, 1904295UR, 1905025UR, 1905065NR, 1905085UR, 1905105UR,  1905135NR, 1905155NR, 1903065WR, 1903185WR, 1903205WR, 1905065WR, 1905105WR, 1905155WR, and 1905225WR;  100310H2HS-V - 1903135UR, 1903155UR, 1903185UR, 1903205UR, 1904015UR, 1904035UR, 1904055UR, 1904085UR, 1904155UR, 1904175UR, 1904175ZR, 1904225UR, 1904245UR, 1904295UR, 1905085WR, and 1905225WR;  100404H2HS-V - 1904175WR, 1904225WR, 1905035WR, 1905065WR, and 1905155WR;  100406H2HS-V - 1903205WR, 1904175WR, and 1905035WR;  100408H2HS-V - 1903085UR, 1903135UR, 1903155UR, 1903155ZR, 1903185UR, 1903225UR,1904015UR, 1904035UR, 1904055UR, 1904225UR, 1903155PR, 1903205WR, 1904105WR, 1904245WR, and 1905065WR;  100410H2HS-V - 1903115UR, 1903155UR, 1903205UR, 1904155UR, 1904175UR, 1904225UR, 1903045WR, and 1905085WR;  100506H2HS-V - 1905155WR;  100508H2HS-V - 1903205WR;  100510H2HS-V - 1903045WR, 1905085WR, and 1905225WR;  100515H2HS-V - 1903065WR, 1903225W6, 1904175WR, 1905085WR, and 1905225WR;  100520H2HS-V - 1904155WR, 1904175WR, 1905085WR, and 1905225WR;  100620H2HS-V - 1903045WR, 1904175WR, 1904225WR, 1905085WR; and   MV-00208HHSA - 1904175YR, 1905065XR, and 1905085YR.
510(k) numbers
["K050954","K091641","K132952","K153594","K161367"]
PMA numbers
not on file
Product code
HCG
Firm FEI
3003523199
Firm city
Aliso Viejo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register