Skip to the content

FDA Medical Device Recalls (openFDA)

Biocare Medical Llc: BioCare Medical Vina Green Chromogen Kit; 25 mL, 100ml IVD. Kits include: Vina Green Chromogen 1. 0 mL; Vina Green Buffer, 25 mL and Vina Green Dropper Bottle. BioCare Medical, Concord, CA 94520. Vina Green is intended for both IHC and ISH applications including HPV, CMV, EBV, Kappa, TTF-1, Ki-67 and other targets/antigens such as blood and lymphatic vessels and basal and myoepithelial cells.

Recall number
Z-1392-2012
Recalling firm
Biocare Medical Llc
Product
BioCare Medical Vina Green Chromogen Kit; 25 mL, 100ml IVD. Kits include: Vina Green Chromogen 1. 0 mL; Vina Green Buffer, 25 mL and Vina Green Dropper Bottle. BioCare Medical, Concord, CA 94520. Vina Green is intended for both IHC and ISH applications including HPV, CMV, EBV, Kappa, TTF-1, Ki-67 and other targets/antigens such as blood and lymphatic vessels and basal and myoepithelial cells.
Status
Terminated
Initiated
2012-03-13
Posted
2012-04-10
Terminated
2012-09-20
Reason
Reports of sporadic failure of specific lots of Vina Green Chromogen Kit, either through crystal formation that obscures positive staining, color not right, or a complete fading of staining by the chromogen.
Root cause
Nonconforming Material/Component
Action
"Medical Device Recall Notification" letters were sent on March 13, 2012 to all direct account. Letters informed customers of the reason for recall and the specific lots and catalog numbers involved, along with the reason for recall, how to identify the recalled product, and the actions to be taken. Customers were to immediately discontinue use of the affected product lot(s). They should dispose of the product according to national, state, and local laws and regulations. The enclosed Return Response Form should be completed and returned. Customers will be credited upon receipt of the Response Form.
Quantity
74 kits
Distribution
Worldwide Distribution -- USA, including the states of WI, KY, CA, GA, IN, MA, IL, MD, NC, LA. AZ, TX, NC, NJ, MO, NY, PA, OH, DC, MI, WA, and CO, and the countries of Canada, Sweden, Germany, Italy, and Switzerland.
Lot, serial or model codes
Catalog number VG807H, lot 060211 and 101211; expiration 5/2013; Catalog number VG807S, lot 053111, 060211; both with expiration of 5/2013; lot 101211, Expiration 8/2013.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NJT
Firm FEI
3004140393
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register