FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc.: V5M TEE Transducer, when used with an ACUSON S2000 ultrasound system that has cardiac options installed and is at software version VA10, VA10A, VA10B, VA10C, VA15, VA15A, VA15B, VA16 or VA16A, manufactured by Siemens Medical Solutions, Mountain View, CA. Multipurpose mobile software controlled diagnostic ultrasound system and on-screen display for thermal and mechanical indices related to bio-effect mechanisms. Its function is to acquire primary or secondary ultrasound echo data. The V5Ms is a multi frequency, multiplane, phased sector array transducer. It can be introduced into the esophagus to obtain images of the heart structure.
- Recall number
- Z-1392-2011
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- V5M TEE Transducer, when used with an ACUSON S2000 ultrasound system that has cardiac options installed and is at software version VA10, VA10A, VA10B, VA10C, VA15, VA15A, VA15B, VA16 or VA16A, manufactured by Siemens Medical Solutions, Mountain View, CA. Multipurpose mobile software controlled diagnostic ultrasound system and on-screen display for thermal and mechanical indices related to bio-effect mechanisms. Its function is to acquire primary or secondary ultrasound echo data. The V5Ms is a multi frequency, multiplane, phased sector array transducer. It can be introduced into the esophagus to obtain images of the heart structure.
- Status
- Terminated
- Initiated
- 2009-08-28
- Posted
- 2011-02-22
- Terminated
- 2011-03-29
- Reason
- If the transducer is not disconnected from the system after an automatic shutdown due to thermal limits being exceeded, then the next time it is used, the thermal safety guards do not take effect. Normally, the thermal limit before automatic image freeze (and therefore power shut off) occurs at 43 degrees Centigrade.
- Root cause
- Software design
- Action
- Siemens sent a Customer Safety Advisory Notification letter on August 28, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that their local Customer Service Engineer would install a software update as soon as it becomes available. Customers were instructed to pass the Customer Safety Advisory Notification letter to all who need to be aware of these issues until the corrective action is completed. For questions regarding this recall call 650-969-9112 or 800-422-8766.
- Quantity
- 97 units
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Canada, China, Czech Republic, Denmark, Germany, Great Britian, Hong Kong, Indonesia, Italy, Japan, Norway, Poland, Portugal, Russian Federation, Singapore, South Korea, Taiwan, and Turkmenistan.
- Lot, serial or model codes
- Model numbers: -10041461: Acuson S2000; -8263703: V5Ms Transducer; -10041524: S2000Cardiac Imaging and Quantification; -10436394: S2000 CV Imaging and Quantification; -10041525: S2000 Advanced Cardiac Analysis; -10041501: S2000 Cardiac Application Module; -10042966: S2000 Cardiac Application Module Upgrade; -10042990: S2000 Cardiac Application Module- No Physio Upgrade; -10042991: S2000 Card Application Module with Physio Upgrade; -100442992: S2000 Card App Module only Upgrade
- 510(k) numbers
- ["K082142"]
- PMA numbers
- not on file
- Product code
- ITX
- Firm FEI
- 2936884
- Firm city
- Mountain View
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28