FDA Medical Device Recalls (openFDA)
Biosite Inc: Triage Cardiac Panel, Biosite Catalog Number 97000HS, 25 individually pouched devices in a labeled kit box. Fluorescence immunoassay kit to be used with the Triage Meter for the quantitative determination of myoglobin, creatine kinase MB fraction (CK-MB), and troponin I in EDTA whole blood and plasma specimens. These three cardiac biomarkers are used as an aid in the diagnosis of myocardial infarction (injury).
- Recall number
- Z-1391-2009
- Recalling firm
- Biosite Inc
- Product
- Triage Cardiac Panel, Biosite Catalog Number 97000HS, 25 individually pouched devices in a labeled kit box. Fluorescence immunoassay kit to be used with the Triage Meter for the quantitative determination of myoglobin, creatine kinase MB fraction (CK-MB), and troponin I in EDTA whole blood and plasma specimens. These three cardiac biomarkers are used as an aid in the diagnosis of myocardial infarction (injury).
- Status
- Terminated
- Initiated
- 2009-04-27
- Posted
- 2009-05-08
- Terminated
- 2010-07-06
- Reason
- The recall was initiated because the use of the affected lot may lead to reporting falsely low values for troponin I and myoglobin, possibly resulting in missed or incorrect diagnosis.
- Root cause
- Process control
- Action
- The recall was initiated with "Urgent Medical Device Recall" notices, dated 4/28/2009, with attached "Customer/Distributor Verification Form" being sent to the affected customers via fax, e-mail or direct mail on 04/28/2009. Customers were informed of the reason for the recall and instructed to immediately discontinue all use of the listed products and discard all affected product in accordance with your their regulations. The customers were give two options; "Option 1: Complete and FAX the enclosed Verification Form within 10 days to confirm your receipt of this notice, to indicate the number of kits discarded from your inventory and product replacement will be sent to you. Option 2: Contact inverness Medical Technical Service at 1-877-441-7440 and select option 2 from the phone prompt. A representative will confirm the number of kits discarded from your inventory and arrange for replacement product to be sent to you." Customers were also provided telephone, fax and e-mail contact information for Biosite should they have any questions about the information contained in the recall notification.
- Quantity
- 401 kits
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Lot Number W44467B
- 510(k) numbers
- ["K030286"]
- PMA numbers
- not on file
- Product code
- MMI
- Firm FEI
- 1000125596
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28