Skip to the content

FDA Medical Device Recalls (openFDA)

Covidien, LP: Plain Gut Absorbable Sutures, PLAIN GUT 5/0 18 P-13 X12 (SG-5686), PLAIN GUT 4-0 18 UD SC-11 DA (SG828-2)

Recall number
Z-1390-2023
Recalling firm
Covidien, LP
Product
Plain Gut Absorbable Sutures, PLAIN GUT 5/0 18 P-13 X12 (SG-5686), PLAIN GUT 4-0 18 UD SC-11 DA (SG828-2)
Status
Open, Classified
Initiated
2023-03-01
Posted
2023-04-12
Terminated
not on file
Reason
Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk.
Root cause
Process control
Action
An "URGENT PRODUCT RECALL" letter dated 3/1/23 was sent to customers. Actions to be Taken: " Immediately identify and quarantine all unused affected product (Refer to Attachment A for identifying affected product). " Please complete and return the enclosed Customer Confirmation Form, even if you have no product to return and email to rs.gmbfcamitg@medtronic.com. " Product purchased directly from Medtronic please contact rs.covidienfeedbackcustomerservice@medtronic.com for Return Goods Authorization (RGA). Your local Medtronic Representative can assist you as necessary in initiating the return of this product. " Credit for the returned affected product will be issued based on the RGA number. " Return all unused affected product(s) in your inventory to: Medtronic, Attn: Field Returns Dept. 195 McDermott Road, North Haven, CT 06473 USA " If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. " Forward this notification to any entities within your organization or to any other organization where the potentially affected product has been transferred or further distributed. Local contact details: Adverse reactions or quality problems experienced with this product should be reported to FDA and Medtronic: " Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or Call FDA (800) FDA-1088 " Contact Medtronic Customer Service Support at rs.covidienfeedbackcustomerservice@medtronic.com or call 800-962-9888, option 2. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic Representative Customer Service at 800-962-9888, option 2.
Quantity
3228 units
Distribution
Worldwide Distribution: US (nationwide) to States of: AR, CA, FL, GA, LA, MA, NC, NY, OH, WA; and countries of: Canada, Colombia, El Salvador, Indonesia, Nepal, Puerto Rico, Taiwan, and Thailand.
Lot, serial or model codes
GTIN: 20884521055756, 20884521750118; Lot: D2G1825FZY, D2H0754FZY, D2E2595ZY
510(k) numbers
["K142656"]
PMA numbers
not on file
Product code
GAL
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register