Skip to the content

FDA Medical Device Recalls (openFDA)

Immuno-Mycologics, Inc: Candida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for use in the ID test, RX only, IVD, CE

Recall number
Z-1389-2025
Recalling firm
Immuno-Mycologics, Inc
Product
Candida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for use in the ID test, RX only, IVD, CE
Status
Open, Classified
Initiated
2025-02-10
Posted
2025-03-18
Terminated
not on file
Reason
Candida Immunodiffusion Antigen may have reduced sensitivity which could result in false negative results.
Root cause
Process change control
Action
On February 10, 2025, Immy issued a "Urgent Medical Device Recall" Notification to affected consignees. Immy asked consignees to take the following actions: 1. Immediately identify, count, and segregate any affected vials you have in your inventory to prevent them from being used or shipped to your customers. As soon as possible, discard your affected inventory. 2. Complete the attached Acknowledgement and Receipt Form (pages 3 and 4 below) even if you do not have any affected stock remaining in your possession. Note: The form is a fillable PDF. You can save it to your computer, fill out electronically and attach to an email. Return the completed form to IMMY using one of the methods below: o Email: customerservice@immy.com; Mail to: Attn: Hunter Conover IMMY, Inc., 2701 Corporate Centre Dr., Norman, OK USA 73069 3. Ensure relevant staff members are informed of this recall, including relevant clinicians. Clinicians should review all negative results. 4. If you have supplied any potentially affected product to another organization, please advise that organization of this recall and send them this notification. Please contact us so we can follow up with them. 5. In case product is in transit, display this letter in a prominent place for one month.
Quantity
30 vials
Distribution
Worldwide - US Nationwide distribution in the states of CA and OK and the countries of Peru France, Austria, Venezuela.
Lot, serial or model codes
Lot #: F4171005/UDI: 00816387020223
510(k) numbers
not on file
PMA numbers
not on file
Product code
GMI
Firm FEI
1627497
Firm city
Norman
Firm state
OK
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register