Skip to the content

FDA Medical Device Recalls (openFDA)

Baxa Corporation: Pharmacy compounding system . It is a twenty-four source compounder for multi-source fluid mixing of both macro and micro ingredient down to 0.2 ml quantities. Product Usage: Pharmacy compounding System for parenteral macro and micro ingredients

Recall number
Z-1389-2012
Recalling firm
Baxa Corporation
Product
Pharmacy compounding system . It is a twenty-four source compounder for multi-source fluid mixing of both macro and micro ingredient down to 0.2 ml quantities. Product Usage: Pharmacy compounding System for parenteral macro and micro ingredients
Status
Terminated
Initiated
2009-03-23
Posted
2012-04-09
Terminated
2012-04-10
Reason
Software issue with compounding pump resulting in over delivery of component ingredients during routine operations.
Root cause
Software design
Action
Baxa sent a Safety Alert letter dated March 23, 2009 to all affected consignees. The letter identified the affected product, problem and recommended actions with specific instructions related to the issue. Additionally, the firm will be providing a warning label for the affected product as soon as possible and is currently working on a product upgrade to correct the issue. For assistance contact Baxa Technical Support at 800-678-2292. The firm is also providing a 24 hour a day, 7 days a week Technical Support for the affected product.
Quantity
1322 units
Distribution
Worldwide Distribution - USA Nationwide including Canada.
Lot, serial or model codes
2 model numbers 2400 D and 2400 M.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NEP
Firm FEI
1419106
Firm city
Englewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register