KaVo Dental Technologies LLC: DEXIS Titanium Intraoral Sensor Product Usage: The DEXIS Titanium Intraoral Sensor is a USB-driven sensor which is intended to acquire dental intraoral radiographic images. The DEXIS / KaVo sensor shall be operated by healthcare professionals, who are educated and competent to perform the acquisition of dental intra-oral radiographs. The DEXIS / KaVo sensor can be used either in combination with special positioning devices to facilitate positioning and alignment with the x-ray beam or it may also be positioned by hand with the assistance of the patient.
- Recall number
- Z-1387-2019
- Recalling firm
- KaVo Dental Technologies LLC
- Product
- DEXIS Titanium Intraoral Sensor Product Usage: The DEXIS Titanium Intraoral Sensor is a USB-driven sensor which is intended to acquire dental intraoral radiographic images. The DEXIS / KaVo sensor shall be operated by healthcare professionals, who are educated and competent to perform the acquisition of dental intra-oral radiographs. The DEXIS / KaVo sensor can be used either in combination with special positioning devices to facilitate positioning and alignment with the x-ray beam or it may also be positioned by hand with the assistance of the patient.
- Status
- Terminated
- Initiated
- 2019-04-29
- Posted
- not on file
- Terminated
- 2020-11-25
- Reason
- The sensor's look up calibration files were incorrectly programmed during manufacturing, which can result in the reduction of image quality.
- Root cause
- Device Design
- Action
- KaVo Dental Technologies, LLC notified customers on about 04/29/2019, via "URGENT: MEDICAL DEVICE RECALL - CORRECTION" letter. Customers were instructed to: 1.) Review the serial number information included with this letter and identify where the affected devices are located in your facility; 2.) For any affected devices that are installed, stop usage and contact Customer Care at 1-888-883-3947 (Option 1) to start the correction via remote connection. See Supplemental Directions on Page 2; 3.) For any affected devices that are unopened or uninstalled, return them to KaVo Dental Technologies, LLC. See Supplemental Directions on Page 2; 4.) If you have sold or forwarded the affected devices to a location outside your control, see Supplemental Directions on Page 2; 5.) Your response is required: Complete and sign the bottom portion of this letter. Return a copy of completed and signed Page 1 to confirm your Recall Responsibility is complete. Note: This can be completed by dental office personnel or on your behalf by your Authorized Dealer. Correction via Reprogramming by a Customer Care Representative: 1.) If you would like to complete your Recall Responsibility by a correction of any affected devices that are already installed, this can be completed free of charge by a Customer Care Representative via remote connection. 2.) The issue can be resolved through a simple and brief software reprogramming activity by replacing the calibration correction file through a remote connection. The sensor hardware meets all specifications, and following the software reprogramming activity, the sensor will operate as intended. 3.) To begin, please contact Customer Care at 1-888-883-3947 (Option 1). 4.) The Customer Care Representative will provide instructions on how to complete the software correction via remote connection. 5.) However, if you prefer to return your affected devices that are installed instead of remote connection, follow the return direction
- Quantity
- 211 devices
- Distribution
- Worldwide Distribution - US Nationwide to AL, AR, AZ, CA, CT, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NV, OH, OK, PA, TN, TX, VA, WA, WI, WV and internationally to Canada.
- Lot, serial or model codes
- Model Number CD63M115.CG; Part Number IOXS-1.012.6180; Serial Numbers: 13095594, 13095595, 13095599, 13095661, 13095665, 13095759, 13095762, 13095766, 13095769, 13095774, 13095780, 13095782, 13095783, 13095784, 13095786, 13095809, 13095810, 13095817, 13095823, 13095827, 13095839, 13095876, 13095890, 13095897, 13095899, 13095900, 13095921, 13095927, 13095929, 13095941, 13095942, 13095955, 13095957, 13096033, 13096034, 13096037, 13096038, 13096040, 13096063, 13096064, 13096065, 13096120, 13096128, 13096145, 13096151, 13096155, 13096162, 13096173, 13096193, 13096195, 13096200, 13096206, 13096212, 13096217, 13096218, 13096220, 13096223, 13096224, 13096227, 13096230, 13096234, 13096235, 13096244, 13096252, 13096257, 13096267, 13096272, 13096286, 13096290, 13096292, 13096295, 13096299, 13096311, 13096318, 13096322, 13096335, 13096340, 13096344, 13096346, 13096348, 13096355, 13096362, 13096383, 13096391, 13096402, 13096405, 13096410, 13096417, 13096426, 13096428, 13096430, 13096437, 13096439, 13096466, 13096479, 13096483, 13096490, 13096491, 13096504, 13096505, 13096518, 13096527, 13096564, 13096574, 13096576, 13096580, 13096583, 13096591, 13096593, 13096602, 13096610, 13096622, 13096725, 13096738, 13096760, 13096804, 13096825, 13096831, 13096883, 13096905, 13096916, 13096928, 13096947, 13096958, 13096960, 13096964, 13096968, 13096970, 13096975, 13096979, 13096982, 13096984, 13096993, 13097017, 13097018, 13097038, 13097039, 13097101, 13097109, 13097118, 13097121, 13097126, 13097153, 13097166, 13097170, 13097181, 13097190, 13097193, 13097195, 13097199, 13097211, 13097214, 13097217, 13097234, 13097240, 13097241, 13097247, 13097249, 13097252, 13097269, 13097287, 13097299, 13097306, 13097309, 13097320, 13097322, 13097326, 13097348, 13097386, 13097389, 13097398, 13097418, 13097419, 13097422, 13097438, 13097439, 13097464, 13097469, 13097476, 13097478, 13097503, 13097517, 13097518, 13097532, 13097546, 13097553, 13097554, 13097557, 13097559, 13097562, 13097580, 13097584, 13097589, 13097603, 13097615, 13097616, 13097619, 13097808, 13097827, 13097838, 13095874, 13095917, 13095930, 13096057, 13096240, 13096448, 13097145, 13097230, 13097245, 13097463, 13097813.
- 510(k) numbers
- ["K172918"]
- PMA numbers
- not on file
- Product code
- MUH
- Firm FEI
- 1017522
- Firm city
- Charlotte
- Firm state
- NC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28