Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Operations, Inc.: Elecsys Troponin I Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human serum and plasma. Elecsys Troponin I is intended to aid in the diagnosis of myocardial infarction.

Recall number
Z-1387-2012
Recalling firm
Roche Diagnostics Operations, Inc.
Product
Elecsys Troponin I Immunoassay For the in-vitro quantitative determination of cardiac troponin I in human serum and plasma. Elecsys Troponin I is intended to aid in the diagnosis of myocardial infarction.
Status
Terminated
Initiated
2012-03-12
Posted
2012-04-11
Terminated
2013-02-06
Reason
An investigation has confirmed a considerably reduced recovery of Li-heparin plasma samples for Troponin I lot 163176 and Troponin I STAT lot 163177. In the worst case, Li-Heparin sample recovery may be as low as 50% of the serum recovery. Correct serum recovery has been confirmed for the affected lots. The issue is caused by the raw material lot (Poly-L-Lysin) instability. Poly-L-Lysin is the
Root cause
Nonconforming Material/Component
Action
The firm, Roche Diagnostics Corporation, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated March 12, 2012 to all customers that received the affected lots of Elecsys Troponin I or Elecsys Troponin I STAT. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of the affected product; discard the affected product from their inventory according to their site's local regulations; if their facility has distributed the affected product to other sites, ensure that this letter is provided to those sites; complete and return the attached fax form via fax to 1-888-912-8457 and file this letter for future references. If you have any questions about the information contained in this letter, please contact Roche Diagnostics Technical Support, 24 hours a day, seven days a week at 1-800-428-2336.
Quantity
739 kits
Distribution
Worldwide distribution. Argentina, Kenya, Austria, South Korea, Belgium, Kuwait, Bahrain, Morocco, Brazil, Mexico, Switzerland, Nigeria, Chile, New Zealand, Columbia, Philippines, Germany, Pakistan, Algeria, Poland, Egypt, Qatar, Spain, Tunisia, France, Turkey, United Kingdom, Taiwan, India, Ukraine, Iran, Venezuela, Italy, South Africa, Jordan, and Armenia.
Lot, serial or model codes
Elecsys Troponin I Immunoassay, 05094810160- lot 163176
510(k) numbers
["K082699"]
PMA numbers
not on file
Product code
MMI
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register