Skip to the content

FDA Medical Device Recalls (openFDA)

Vyaire Medical: enFlow Disposable Cartridge with IV Extension Set, used for intravenous warming therapy with fluid and blood solutions.

Recall number
Z-1386-2019
Recalling firm
Vyaire Medical
Product
enFlow Disposable Cartridge with IV Extension Set, used for intravenous warming therapy with fluid and blood solutions.
Status
Terminated
Initiated
2019-03-13
Posted
2019-06-26
Terminated
2020-09-22
Reason
Testing has demonstrated aluminum elution from the enFlow Disposable Cartridge during fluid warming.
Root cause
Device Design
Action
Vyaire Medical notified consignees on about 03/13/2019 via URGENT NOTIFICATION OF RECALL letter. Consignees were instructed to conduct a 100% physical inventory to identify any affected product on hand, destroy all affected product in-stock in accordance with each facility's destruction protocol and if a facility is unable to destroy the product on site, or requires assistance, they were instructed to contact VyaireSupport@stericycle.com or call for assistance. Customers were also instructed to complete and return the Customer Response Form to VyaireSupport@stericycle.com, report any adverse reactions or quality problems to Vyaire's Technical Support Department by e-mail GMB-RS-DISP-Complaint-Intake@carefusion.com or telephone at 1-800-231-2466 prompt #1. If customers further distributed the product, they were instructed to forward a copy of the URGENT NOTIFICATION OF RECALL letter and Response Form to all customers and include the contact information to Vyaire for tracking purposes.
Quantity
5,782,820 units total
Distribution
Worldwide Distribution. US nationwide, Brazil, Canada, Chile, Colombia, Ecuador, Hong Kong, Hungary, Israel, India, Jordan, Korea, Kuwait, Kazakhstan, Lebanon, Macedonia, Mexico, Malaysia , Myanmar, Netherlands, New Zealand, Oman, Peru, Qatar, Russian Federation, Saudi Arabia, Thailand, Turkey, United Republic of Tanzania, South Africa, Panama, and Viet Nam.
Lot, serial or model codes
Part Number: 980202EU. All units distributed through March 7, 2019.
510(k) numbers
["K060537","K130867"]
PMA numbers
not on file
Product code
LGZ
Firm FEI
2050001
Firm city
Mettawa
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register