Radiometer America Inc: Blood Gas and Co-Ox2 Electrolyte & Metabolyte Analyzers, ABL-700 Series, manufactured from 8/1998 through 11/2004.

Recall number
Z-1386-06
Recalling firm
Radiometer America Inc
Product
Blood Gas and Co-Ox2 Electrolyte & Metabolyte Analyzers, ABL-700 Series, manufactured from 8/1998 through 11/2004.
Status
Terminated
Initiated
2005-09-29
Posted
2006-08-10
Terminated
2012-04-17
Reason
Under certain circumstances, the ABL 700 & 800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and Sodium (Na).
Root cause
Other
Action
The consignees were notified of this recall by a letter, dated 9/29/2005. The letter informs the customers of this potential problem and the availability of new updated software which will correct the problem. The letter goes on to inform the customer that a Radiometer Service Engineer wil be contacting them shortly in order to arrange for a convenient time to install the software update. Along with the letter, the customer is provided a copy of an additional page and a CD concerning the function and operation of the new software, which is to be added to the customer''s Operator''s Manual.
Quantity
1615 devices
Distribution
Nationwide.
Lot, serial or model codes
All analyzers manufactured from 8/1998-11/2004, are subject to recall. The devices contain serial numbers 902-441RxxxNxxx. The Rxxx number denotes manufacturing run, and the Nxxx number denotes the individual instrument within the run.
510(k) numbers
["K002290","K980130","K991417","K992859"]
PMA numbers
not on file
Product code
JFP
Firm FEI
1523456
Firm city
Westlake
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register