Sendx Medical Inc: SenDx Medical Inc ABL80FLEX Blood Gas, pH, Electrolyte, and Metabolite System The ABL80 FLEX is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, and hematocrit in whole blood. The ABL80 FLEX system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.

Recall number
Z-1384-2012
Recalling firm
Sendx Medical Inc
Product
SenDx Medical Inc ABL80FLEX Blood Gas, pH, Electrolyte, and Metabolite System The ABL80 FLEX is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, and hematocrit in whole blood. The ABL80 FLEX system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.
Status
Terminated
Initiated
2012-01-30
Posted
2012-04-09
Terminated
2012-07-13
Reason
The product can cause the analyzer to report values outside the published performance specifications for pH, pCO2, cNa+, cK+, cCa2+ and cCl-.
Root cause
Nonconforming Material/Component
Action
SenDx Medical sent Customers notification letters dated January 30th, 2012, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customer. We have recently become aware of a possible condition in a limited number of ABLBO FLEX analyzers configured with a specific Control PCe lerinted Circuit Board), which can cause the analyzer to report values outside the published performance specifications for pH, pCOz, cNa*, cK*, cCa2* and cCl-. Customers were instructed to immediately implement Alternative 1 or 2 until the signal data PCB is replaced. Alterative I: Stop the use of all affected analyzers until corrective actions can be performed on these analyzers. Alternative 2: Verify proper function of the affected analyzers immediately following every patient sample measurement by measuring a manual QC sample. Customers were asked to complete the fax form acknowledging receipt of the notice and provided a contact for Technical Support at 1-800-736-0600 opt.4.
Quantity
819 analyzers and 44 signal data PCBs
Distribution
One US consignee in state of OH. Other consignees are international such as Australia, France, Netherlands, Denmark, China, United Kingdom, Turkey, Switzerland, Japan, India, Poland, Spain, and Germany. Distribution list submitted with supporting doc.s
Lot, serial or model codes
Model number: 393-839 Serial number(s): 304157, 304160 and 304162 to 305414 for ABL80 FLEX analyzers
510(k) numbers
["K051804"]
PMA numbers
not on file
Product code
CHL
Firm FEI
2027541
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register