Covidien, LP: Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT

Recall number
Z-1383-2021
Recalling firm
Covidien, LP
Product
Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT
Status
Terminated
Initiated
2021-03-04
Posted
not on file
Terminated
2024-07-12
Reason
Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.
Root cause
Under Investigation by firm
Action
Covidien/Medtronic issued Urgent Medical Device Recall via Federal Express and/or certified mail beginning March 4, 2021, states reason for recall, health risk and action to take: 1. Please immediately quarantine and discontinue use of affected item codes with associated lot numbers listed on Attachment A. 2. Please return affected product as indicated below. All unused products from the affected item codes and associated lot numbers must be returned. 3. If you have distributed the Covidien Signia" small diameter curved tip intelligent reload listed on Attachment A, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Quantity
5928 units WW: 4212 US 1716 OUS
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.
Lot, serial or model codes
Lot Numbers: N0A1038UY N0E0016UY N0F0616UY N0K0934UY N0B0493UY N0F0254UY N0F0817UY N0M0334UY N0C0436UY N0F0518UY N0G0388UY N9K1095UY N0D0731UY N0F0615UY N0J0859UY N9L0797UY  UDI: 10884521741843 20884521741840
510(k) numbers
["K191070"]
PMA numbers
not on file
Product code
GDW
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register