FDA Medical Device Recalls (openFDA)
Synthes (USA) Products LLC: PRODISC Orthopedic Manual Surgical Instrument, strut used in spinal surgery. Part numbers 03.820.350 and 03.820.352. Used to hold the disc segment open and facilitate discectomy
- Recall number
- Z-1383-2017
- Recalling firm
- Synthes (USA) Products LLC
- Product
- PRODISC Orthopedic Manual Surgical Instrument, strut used in spinal surgery. Part numbers 03.820.350 and 03.820.352. Used to hold the disc segment open and facilitate discectomy
- Status
- Terminated
- Initiated
- 2017-01-31
- Posted
- 2017-02-27
- Terminated
- 2017-12-01
- Reason
- These items are incorrectly etched and labeled. The item etched and labeled as 10mm Strut 03.820.350 lot 5919005 is a 12 mm Strut. The item etched and labeled as 12 mm Strut 03.820.352 lot 5919006 is a 10 mm Strut.
- Root cause
- Under Investigation by firm
- Action
- DePuy Synthes mailed an Urgent Notice to customers affected by the issue on 02/13/2017. The reason for the recall was explained, risks associated with the issue were identified, and actions to be taken were addressed. Customers were asked to complete the enclosed Verification form and mail it back to obtain a Return Authorization Number for the product's return. Customers with questions were instructed to call 610-719-5450 or contact their DePuy Synthes Sales Consultant. For questions regarding this recall call 800-479-6329.
- Quantity
- 2 units
- Distribution
- US distribution to TX and LA
- Lot, serial or model codes
- Part numbers 03.820.350 Lot 5919005 and 03.820.352, lot 5919006
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 3008812560
- Firm city
- West Chester
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28