Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens Axiom Luminos dRF solid state x-ray imager Solid state x-ray imager (flat panel/dig. imager)

Recall number
Z-1383-2012
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens Axiom Luminos dRF solid state x-ray imager Solid state x-ray imager (flat panel/dig. imager)
Status
Terminated
Initiated
2012-02-16
Posted
2012-04-09
Terminated
2014-11-06
Reason
During regular product monitoring, Siemens became aware of a potential issue with the AXIOM Luminos dRF. Firm became aware of a potential issue with torque wrenches - during a regular scheduled calibration of the wrench, it was discovered that the wrench was out of the specification. The wrench was over-torquing approximately 21% for low-end readings (50 ftlbs) and 9% for high-end readings (250
Root cause
Nonconforming Material/Component
Action
Siemens initiated a recall/correction on February 16, 2012 by issuing Update Instruction XP056/11/S to the affected customers. All screws on the unit support on AXIOM Luminos dRF units will be replaced by a Siemens service engineer. For information regarding this recall call 610-219-4834.
Quantity
2
Distribution
Nationwide Distribution including California
Lot, serial or model codes
Model number 10094200, serial numbers 3019 and 1358
510(k) numbers
["K062623"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register