FDA Medical Device Recalls (openFDA)
Intuitive Surgical, Inc.: da Vinci S Surgical System IS2000, Camera Arm Drape, 20 Pack. Microtek Medical Equipment Drapes are to be used to cover a variety of surgical and non-surgical equipment in various settings throughout the clinical setting.
- Recall number
- Z-1381-2015
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- da Vinci S Surgical System IS2000, Camera Arm Drape, 20 Pack. Microtek Medical Equipment Drapes are to be used to cover a variety of surgical and non-surgical equipment in various settings throughout the clinical setting.
- Status
- Terminated
- Initiated
- 2015-03-16
- Posted
- 2015-04-06
- Terminated
- 2015-12-03
- Reason
- Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to tear.
- Root cause
- Packaging process control
- Action
- Firm issued field safety notice to all customers on March 16, 2015. Field safety notice includes acknowledgement form to be returned to recalling firm. A second letter was sent on May 12, 2015 to inform users that the date that the drape product unaffected by the issue will not be availalble untisl October/November 2015, not May 2015 as stated in the original communication.
- Quantity
- 442,475 total all drape models
- Distribution
- Worldwide Distribution-Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China (including Hong Kong), Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, France, Germany, Greece, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Malaysia, Mauritius, Mexico, Monaco, Netherlands, New Zealand, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, Vietnam.
- Lot, serial or model codes
- Product number 400016-04; da Vinci (Standard System) Camera Arm Drape, 20 Pack. Lot Numbers for Cloudy/Waxy issue: ALL lot numbers Lot Numbers for Tear issue: None
- 510(k) numbers
- ["K050322"]
- PMA numbers
- not on file
- Product code
- KKX
- Firm FEI
- 3001675293
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28