Skip to the content

FDA Medical Device Recalls (openFDA)

International Technidyne Corp.: Hemochron Jr. Whole Blood Microcoagulation system Low Range Activated Clotting Time. ACT-LR Cuvettes. In vitro diagnostic use, non-sterile, Packaged in individual foil pouches. 45 foil pouches per shelf carton box. Mfg. by International Technidyne Corp, Edison, NJ 08820.

Recall number
Z-1381-05
Recalling firm
International Technidyne Corp.
Product
Hemochron Jr. Whole Blood Microcoagulation system Low Range Activated Clotting Time. ACT-LR Cuvettes. In vitro diagnostic use, non-sterile, Packaged in individual foil pouches. 45 foil pouches per shelf carton box. Mfg. by International Technidyne Corp, Edison, NJ 08820.
Status
Terminated
Initiated
2005-04-27
Posted
2005-08-23
Terminated
2005-08-25
Reason
Validation had been initiated for a molded component of the finished device but not completed. Also, the device did not meet mean clotting time using Normal Donor Blood as media.
Root cause
Other
Action
Recall letter was sent out on 4/27/2005 via First Class Mail.
Quantity
924 boxes x 45 cuvettes per box
Distribution
Products were distributed domestically to customers nationwide. Products were also distributed internationally to customers in the following countries: Italy, Sweden, Norway, Germany, and Japan. Products were distributed to the VA Medical Center, Los Angeles, CA 90073.
Lot, serial or model codes
Catalog No. JACT-LR Lot No''s:  M4JLR164, Exp. 3/2006; M4JLR165, Exp. 3/2006; A5JLR001, Exp. 4/2006; A5JLR002, Exp. 4/2006; A5JLR003, Exp. 4/2006; C5JLR032, Exp. 6/2006.
510(k) numbers
["K960749"]
PMA numbers
not on file
Product code
JBP
Firm FEI
1000526865
Firm city
Edison
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register