Skip to the content

FDA Medical Device Recalls (openFDA)

Waldemar Link GmbH & Co. KG (Mfg Site): Tibial Component Modular. orthopedic prosthesis. Model Nos: Small, W 16-2817/32 Medium, W 16-2817/35 Large, W 16-2817/37

Recall number
Z-1380-2022
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Product
Tibial Component Modular. orthopedic prosthesis. Model Nos: Small, W 16-2817/32 Medium, W 16-2817/35 Large, W 16-2817/37
Status
Open, Classified
Initiated
2022-05-24
Posted
not on file
Terminated
not on file
Reason
There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
Root cause
Nonconforming Material/Component
Action
On May 24, 2022, the firm notified affected customers via Urgent Field Safety Notice email. Customers were provided a modified surgical technique providing information for the cementing technique for Tilastan proximal spacers to minimize the risk. No product return is required.
Quantity
3785
Distribution
Domestic distribution to AL, CA, FL, GA, IL, IN, KS, MI, MN, NV, NY, OH, TN, TX, WA. Foreign distribution to Algeria Argentina Australia Austria Belgium Bulgaria Canada China Colombia Denmark Ecuador Estonia Finland France Germany Great Britain Greece Hungary India Indonesia Israel Italy Latvia Libya Lithuania Mexico Netherlands Norway Pakistan Peru Philippines Poland Romania Saudi Arabia Slovakia Slovenia Spain Sudan Sweden Switzerland Thailand Turkey Ukraine United Arab Emirates Uruguy Vietnam
Lot, serial or model codes
UDI-DI:  04026575370870 Small, W; 04026575370887 Medium, W; 04026575370894 Large, W;  Serial/Lot Numbers: 210308/0020 210308/0035 210222/0789 151116/4741
510(k) numbers
["K212742"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3003386935
Firm city
Norderstedt
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register