Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Baxter Supple Peri-Guard Repair Patch (legacy product codes): a) PC0404SNBIO, b) PC0608NSBIO, c) PC0814SNBIO, d) PC1016SNBIO

Recall number
Z-1378-2024
Recalling firm
Baxter Healthcare Corporation
Product
Baxter Supple Peri-Guard Repair Patch (legacy product codes): a) PC0404SNBIO, b) PC0608NSBIO, c) PC0814SNBIO, d) PC1016SNBIO
Status
Open, Classified
Initiated
2024-02-28
Posted
2024-03-26
Terminated
not on file
Reason
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
Root cause
Labeling Change Control
Action
Baxter issued a "SAFETY ALERT" to its consignees on 02/28/2024 via USPS first class mail. the notice explained the problem, potential hazard involved with off-label use of the product and the actions to be taken. The firm requested that the notice be forwarded all affected facilities or departments with institutions. Distributors are directed to notify their customers. For questions, contact Baxter Healthcare Center for Service at 888-229-0001, between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday.
Quantity
9549 units
Distribution
US Distribution to states of: CT, VA, TX, PA and IL; and OUS (foreign) countries of: Austria, Finland, Germany, Greece, Ireland, Italy, Poland, Spain, Sweden, Switzerland, and United Kingdom.
Lot, serial or model codes
a) PC0404SNBIO, UDI/DI 00085412530642; b) PC0608SNBIO, UDI/DI 00085412530673; c) PC0814SNBIO, UDI/DI 00085412530697; d) PC1016SNBIO, UDI/DI 00085412530710; ALL LOTS WITHIN EXPIRY
510(k) numbers
["K983162"]
PMA numbers
not on file
Product code
FTM
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register