FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corporation: Baxter Supple Peri-Guard Repair Patch (legacy product codes): a) PC0404SNBIO, b) PC0608NSBIO, c) PC0814SNBIO, d) PC1016SNBIO
- Recall number
- Z-1378-2024
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter Supple Peri-Guard Repair Patch (legacy product codes): a) PC0404SNBIO, b) PC0608NSBIO, c) PC0814SNBIO, d) PC1016SNBIO
- Status
- Open, Classified
- Initiated
- 2024-02-28
- Posted
- 2024-03-26
- Terminated
- not on file
- Reason
- A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
- Root cause
- Labeling Change Control
- Action
- Baxter issued a "SAFETY ALERT" to its consignees on 02/28/2024 via USPS first class mail. the notice explained the problem, potential hazard involved with off-label use of the product and the actions to be taken. The firm requested that the notice be forwarded all affected facilities or departments with institutions. Distributors are directed to notify their customers. For questions, contact Baxter Healthcare Center for Service at 888-229-0001, between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday.
- Quantity
- 9549 units
- Distribution
- US Distribution to states of: CT, VA, TX, PA and IL; and OUS (foreign) countries of: Austria, Finland, Germany, Greece, Ireland, Italy, Poland, Spain, Sweden, Switzerland, and United Kingdom.
- Lot, serial or model codes
- a) PC0404SNBIO, UDI/DI 00085412530642; b) PC0608SNBIO, UDI/DI 00085412530673; c) PC0814SNBIO, UDI/DI 00085412530697; d) PC1016SNBIO, UDI/DI 00085412530710; ALL LOTS WITHIN EXPIRY
- 510(k) numbers
- ["K983162"]
- PMA numbers
- not on file
- Product code
- FTM
- Firm FEI
- 1417572
- Firm city
- Deerfield
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28